14. Read the marker value from the analyzer display and add 5 dE% Write the
16. Read the marker value from the analyzer display and add 5 dB. Write the
calculated value on the test record.
17. Repeat steps 10 through 16 but enter the power value listed in row Pa and
add 10 dB to arrive at the calculated value.
18. Repeat steps 10 through 16 but enter the power value listed in row P3 and
subtract 6 dB to arrive at the calculated value.
19. Repeat steps 10 through 16 but enter the power value listed in row P4 and
subtract 10 dB to arrive at the calculated value.
20. Press [StopJ (20) (ZJJ.
22. Refer to 'hble 2-4 and enter the power value for P~ef for the particular
analyzer under test. After you enter the value, press (iiJ
23. Wait for the analyzer to complete a full sweep.
24. Press ~Dirplayj
27. Enter the power value that is listed in the PI row of l?&le 2-4 for the
particular analyzer under test. After you enter the value, press (xl.
28. Wait for the analyzer to complete a full sweep.
29. Press (pixxg
30. Read the marker value from the analyzer display and add 5 dB. Write the
calculated value on the test record.
.
.