19. Adjust the mid band power offset resistor (A59 R71) for a power meter
reading that is equal to the value (f0.2 dB) listed under "Power B" in
20. Press=
21. A@st the mid band power slope resistor (A59 R88) for a power meter
reading that is equal to the value (&2 dl3) listed under "Power B" in
23.
Adjust the low band power offset resistor (A59 R70) for a power meter
reading that is equal to the value listed under "Power B" in !lhble 34.
24. Press(MenuJ
25. Adjust the low band power slope resistor (A59 R86) for a power meter
reading that is equal to the value listed under "Power B" in 'Ihble 3-4.
27.
Adjust the high band power offset resistor (A59 R80) for a power meter
reading that is equal to the value listed under "Power B" in Table 3.4.
29.
Adjust the high band power slope resistor (A59 RN) for a power meter
reading that is equal to the value listed under "Power B" (f4 d.B) in
32. Enter the power value from the "Power C" column in lhble 3.4 for the
particular analyzer that you are adjusting.