Agilent Technologies HP 8719D Service Manual page 62

Network analyzers
Table of Contents

Advertisement

20
A i r l i n e
S t e p p e d i m p e d a n c e airline
0.6
(rho -
42.
When all measurements are complete, the m$J@ softkey menu will appear.
Disconnect the verification device.
43.
Enter Test 28 (using step keys, entry keys, or front panel knob). Repeat
steps 38 through 42 with the 40 or 50 dl3 attenuator.
44.
Enter 'It% 29 (using step keys, entry keys, or front panel knob). Repeat
steps 38 through 42 with the 5OU airline. For an example of how to perform
proper airline connections, refer to the following @urea
Be especially careful not to drop either the center conductor or
the outer conductor when handling these airlines. Irreparable
damage will result if these devices are dropped.
conductor of the test port with the center conductor of the
airline. Ground yourself to prevent electrostatic discharge
(ESD).
X
X
X
X
X
X
Magnitude
X
X
X
X
X
X
X
X
System Verification and