accesses softkeys that affect the way tests (routines)
run, or supply necessary additional data.
toggles the repeat function on and off. When the
function is ON, the selected test will run 10,000
times unless you press any key to stop it. The
analyzer shows the current number of passes and
fails.
toggles the record function on and off. When the
function is ON, certain test results are sent to
a printer via HP-IB. This is especially useful for
correction constants. The instrument must be
in system controller mode or pass contro1 mode
to print (refer to the 'Printing, Plotting, and
Saving Measurement Results" chapter in the HP
selects either NORMal or SPeCiaL (tighter) limits for
the Operator's Check. The SPCL limits are useful for
a guard band.
accesses the following Edit List menu to allow
modification of the external power loss data table.
Service Key Menus and Error Messages
10-7