loads calibration.
19. When the measurement is complete, press ~~~~~. Leave the load
connected to the reference test port 1 cable.
20. When the analyzer finishes measuring the standard, connect the open that is
supplied in the calibration kit to reference test port 2.
22. When the analyzer linishes measuring the standard, connect the short that
is supplied in the calibration kit to reference test port 2.
23. Press
24. When the analyzer -es measuring the standard, connect the 60 ohm
termination to reference test port 2.
25. Press
loads calibration.
27. When the measurement is complete, press
Leave the load connected to the reference test port 2 cable.
28. 'I'he analyzer briefly displays COHPUTIBG CAL COEFFICIENTS.
30. Connect the two test port return cables together to form a "thru"
configuration, as shown in Figure 2-10.
epending on which device is used. If
depending on which device is used. If
easure a lowband load to complete the
System VMiwtion and
Performance Tests
2-23