7.
Observe the analyzer for the results of the routine:
If the analyzer displays the message Serial Cor DOME, you have completed
this procedure
8. If no more correction constant routines are going to be performed, return the
1. If the analyzer does not display DOlIE, then either the serial number that
you entered in steps 3 and 4 did not match the required format or a serial
number was already stored. Check the serial number recognized by the
analyzer:
b. Look for the serial number displayed on the analyzer screen.
c. Rerun this adjustment test if the serial number is not displayed.
2.
If the analyzer continues to fail this adjustment routine, contact your nearest
Hewlett-Packard sales and service oflice.