ADC Offset Correction Constants (Bst 48)
These correction constants improve the dynamic accuracy by shifting small
1. Make sure the A7 jumper/switch is in the Alter position (AU').
Note
3.
Observe the analyzer for the results of the adjustment routine:
If the analyzer displays ADC Ofs Cor DONE, you have completed this
procedure
4.
If no more correction constant routines are going to be performed, return the
5. Perform the "EEPROM Backup Disk Procedure" located on page 3-34.
In Case of Diffhlty
If the analyzer displays ADC Ofs Cor FAIL, refer to the 'Digital Control
Troubleshooting" chapter.