ESD (electrostatic discharge) grounding wrist strap
n
Removal
1. Disconnect the power cord.
2. Remove the four bottom feet and bottom cover (refer to "Covers" in this
chapter).
3. Remove the four screws (item 1) on the rear panel.
4. Turn the analyzer over and remove the screw (item 2) that secures the CPU
board to the deck.
5. Disconnect the ribbon cable (item 3), sliding your Enger nail between the
cable and the connector.
6. Disconnect the ribbon cable (item 4) from the CPU board.
7. Slide the board towards the front of the instrument so that it disconnects
from the three standoffs (item 5).
8. Disconnect the ribbon attached at the rear of the CPU board.
9.
Lift the board off of the standoffs.
Replacement
1. Reverse the order of the removal procedure.
14-22