4. Dynamic Range Performance 'I&t
Dynamic range is checked by comparing the noise floor to the test port output
power level.
Required Equipment and 'lbols (HP 8719D/2OD Only)
Calibration Kit.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .HP 85052B/D
RF Cable Set.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .HP 85131C/D
Calibration Kit.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .HP 85056B/D
RF Cable Set.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . , . HP 85133C/D
Note
Other calibration kits may be used if you are working in a
different connector type.
1. Tb preset the analyzer, press LprecetJ
2. To enter the values for the first frequency segment:
a.
below, press:
b.
greater, press:
2-44 System Veriicatien and