NOT ALLOWED DURING POWER METER CAL
Error Number When the analyzer is performing a power meter calibration, the
HP-IB bus is unavailable for other functions such as printing or
198
plotting.
OVERLOAD ON INPUT A, POWER REDUCED
Error Number You have exceeded approximately + 14 dBm at one of the test
ports. The RF output power is automatically reduced to
58
-85 dBm. The annotation Pl.l appears in the left margin of the
display to indicate that the power trip function has been
activated. When this occurs, reset the power to a lower level,
then toggle the f%XlRCE PWR on QFF softkey to switch on the
power again.
OVERLOAD ON INPUT B, POWER REDUCED
Error Number You have exceeded approximately + 14 dBm at one of the test
59
ports, The RF output power is automatically reduced to
-85 dBm. The annotation P.U appears in the left margin of the
display to indicate that the power trip function has been
activated. When this occurs, reset the power to a lower level,
then toggle the ~~~::~~~,.~~ @F softkey to switch on the
power again.
1048
Service Key Menus and Error Messages