31. Press
32. Read the marker value from the analyzer display and add 5 dB. Write the
calculated value on the test record.
33. Press m (20) (ZJJ.
34. PressLstoe](ZJ~.
35. Repeat steps 21 through 32.
36. Press m Iso] (EJJ.
37. Repeat steps 21 through 26.
38. Enter the power value that is listed in the P2 row of Table 2-4 for the
particular analyzer under test. After you enter the value, press @.
39. Wait for the analyzer to complete a full sweep.
41. Read the marker value from the analyzer display and add 10 dB. Write the
43. Read the marker value from the analyzer display and add 10 dB. Write the
calculated value on the test record.
44. Press(3ijYJ@iTJ(@J
45. Repeat steps 21 through 26.
46. Enter the power value that is listed in the P3 row for the particular
analyzer under test. After you enter the value, press (xl.
47. Wait for the analyzer to complete a full sweep.
49. Read the marker value from the analyzer display and subtract 5 dB. Record
the calculated value on the test record.
50. Press
51. Read the marker value from the analyzer display and subtract 5 dB. Record
the calculated value on the test record.
242 System Verification and
e test record.
.
.
.