When the prompt, Set source to 2.345 GHz, then continue appears, use
3.
the front panel knob to adjust the fr
5 MHz of 2.345 GHz. Press
For the HP 8722D, when the prompt, Set source to 19.550 GHz, then
4.
continue appears, use the front panel knob to adjust
analyzer source to within 5 MHz of 19.550 GHz. Press
frequency is set.
5.
When Pretune Adj DOIiE is displayed, press w
If no more correction constant routines are going to be performed, return the
6.
Perform the "EEPROM Hackup Disk Procedure' located on page 3-34.
7.
In Case of DifRculty
If any error messages appear, refer to "Source Troubleshooting."
when the frequency is set.
Adjustments and Correction Constants