Agilent Technologies E5071C Service Manual

Agilent Technologies E5071C Service Manual

Ena rf network analyzers
Hide thumbs Also See for E5071C:
Table of Contents

Advertisement

Quick Links

Agilent E5071C ENA Series RF Network Analyzers
Service Guide
Third Edition
Manufacturing No. E5071-90120
October 2007

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies E5071C

  • Page 1 Agilent E5071C ENA Series RF Network Analyzers Service Guide Third Edition Manufacturing No. E5071-90120 October 2007...
  • Page 2 WARNINGS elsewhere in this manual may impair the protection provided by the equipment. In addition it violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements. NOTE E5071C comply with INSTALLATION CATEGORY II and POLLUTION DEGREE 2 in IEC61010-1.
  • Page 3 Because of the danger of introducing additional hazards, do not install substitute parts or perform unauthorized modifications to the instrument. Return the instrument to a Agilent Technologies Sales and Service Office for service and repair to ensure that safety features are maintained.
  • Page 4 Safety Symbol General definitions of safety symbols used on the instrument or in manuals are listed below. Instruction Manual symbol: the product is marked with this symbol when it is necessary for the user to refer to the instrument manual. Alternating current.
  • Page 5 Certification Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institution’s calibration facility, or to the calibration facilities of other International Standards Organization members.
  • Page 6 Typeface Conventions Sample (bold) Boldface type is used when a term is defined or emphasised. Sample (Italic) Italic type is used for emphasis. Indicates a hardkey (key on the front panel or key / [Sample] key external keyboard) labeled “Sample.” “key” may be omitted.
  • Page 7: Table Of Contents

    E5071C Performance Test Calculation Sheet ........
  • Page 8 How to exit from the E5071C Measurement View ....... . .
  • Page 9 Contents To Check the LCD..............199 To Check the External Keyboard .
  • Page 10 Contents LCD Assembly ..............253 Inverter Assembly.
  • Page 11 Contents Handler I/O Board Replacement............286 Tools Required .
  • Page 12 Contents Tools Required ..............310 Removal Procedure.
  • Page 13 Location of the latest E5071C Firmware ........
  • Page 14 Contents...
  • Page 15: General Information

    General Information The Service Manual is a guide to servicing the E5071C ENA Series Network Analyzer. The manual contains information requisite to do performance tests, adjustments, troubleshooting, and repairs.
  • Page 16: Precautions

    Do not allow any computer virus to infect the system. This machine has no virus check function nor anti-virus software installed. Agilent Technologies will not be held liable for any failure or damage arising from negligence regarding these prohibitions and warnings.
  • Page 17: Organization Of Service Manual

    This appendix contains the information required to adapt this manual to versions or configurations of the E5071C manufactured earlier than the current printing date of this manual. The information in this manual applies directly to E5071C units with the serial number that is printed on the title page of this manual.
  • Page 18 Appendix D , “Power Requirement,” Appendix E, “Messages,” The E5071C can display error messages as well as messages that indicate the internal operating status of the equipment. This appendix explains what these messages mean by listing them in alphabetical order.
  • Page 19: Instrument Covered By This Manual

    The supplement for this manual is identified by this manual's printing data and is available from Agilent Technologies. If the serial prefix or number of an instrument is lower than that on the title page of this manual, see Appendix A, Manual Changes. For information concerning, a serial number prefix that is not listed on the title page or in the Manual change supplement, contact the nearest Agilent Technologies office.
  • Page 20: Required Equipment

    General Information Required Equipment Required Equipment Table 1-1 lists the recommended equipment for performing maintenance on the E5071C. Table 1-1 Recommended Test Equipment Equipment Critical specifications Recommended Model Qty. Frequency Counter Frequency: 50 MHz to 8.5 GHz Agilent 53181A with Opt.010 &...
  • Page 21 General Information Required Equipment Table 1-1 Recommended Test Equipment Equipment Critical specifications Recommended Model Qty. Cable 20 inch 50 ohm cable Agilent p/n 5062-6691 Adapter N(m)-BNC(f) Adapter Agilent p/n 1250-0780 Torque Wrench Size: 3/4 inch Agilent p/n 8710-1766 Torque: 136 N-cm Personal Computer Windows XP, VEE7.5 or later with GPIB board...
  • Page 22: Power Meter Accuracy Test

    Power Meter Accuracy Test Power Meter Accuracy Test This test is intended for power meters used in testing the E5071C. The “Dynamic Accuracy Test” requires the use of a power meter that has been calibrated to a higher accuracy than the standard power meter.
  • Page 23 General Information Power Meter Accuracy Test NOTE It is recommended that a copy of the test record on page 26 be made, and the values be recorded on the copy, thus preserving the original for future use. Test Procedure NOTE This procedure assumes the use of the recommended equipment model numbers.
  • Page 24 General Information Power Meter Accuracy Test • POLARITY: NORMAL • RANGE: 1 mW • FUNCTION: CALIBRATE Step 6. Allow the equipment to warm up for approximately 30 minutes. Do not change any connections or control setting during this time. Step 7. Zero and calibrate the power meter channel to which the range calibrator is connected: a.
  • Page 25 General Information Power Meter Accuracy Test Step 17. Record the DVM voltage reading as value D in the test record. Step 18. Wait for the power meter reading to settle (no settling drift within 20 seconds). Step 19. Record the power meter reading as value E in the test record. Step 20.
  • Page 26: Test Record For Power Meter Accuracy Test

    General Information Power Meter Accuracy Test TEST RECORD FOR POWER METER ACCURACY TEST Power Meter Tested Model Number: Test Date: Serial Number: Tested by: Test Equipment Used Range Calibrator: Model No.: Serial No.: Digital Voltmeter: Model No.: Serial No.: Test Results Range Channel A Channel B...
  • Page 27: Performance Test

    Performance Test This chapter provides information on how to verify the E5071C performance.
  • Page 28: Introduction

    Performance Test Introduction Introduction This literature provides the performance test procedures for the Agilent E5071C RF Network Analyzers. The performance test names are listed in Table 2-1. The test procedures are described sequentially in the following pages. The test name indicates the tested performance and to which performance group the tested performance belongs.
  • Page 29: Test Equipment Required

    Softkey Selection Procedure for Performance Test NOTE The procedure of the softkey selection depends on the firmware revision. Please note that some softkey selection procedures for your E5071C might be a little bit different from the procedure described in this chapter. Chapter 2...
  • Page 30: Aux Input Test

    Performance Test 1. AUX INPUT TEST 1. AUX INPUT TEST Description This test checks the measurement accuracy of AUX Input. This test is checked, compared with the measured value of Multimeter by inputting the DC voltage to AUX-1 and 2 ports from function generator.
  • Page 31: Frequency Accuracy Test

    2. FREQUENCY ACCURACY TEST Description This test checks the frequency accuracy of the E5071C test port output signal. The frequency accuracy is checked at 50 MHz and 8.5 GHz with a frequency counter. Since the E5071C employs a PLL frequency synthesizer for the signal source, the frequency accuracy test at these two frequency points can verify the accuracy for the entire frequency range.
  • Page 32: Rf Output Level Accuracy And Flatness Test

    TEST Description This test checks the level accuracy and frequency flatness of the E5071C test port output signal. The level accuracy is checked for an output power level setting of 0 dBm at 50 MHz using a power meter. The frequency flatness is tested by measuring the power level at 26 frequency points from 9 kHz to 8.5 GHz and calculating the differences of the power meter...
  • Page 33: Rf Output Level Linearity Test

    4. RF OUTPUT LEVEL LINEARITY TEST Description This test checks the level accuracy of the E5071C test port output signal across the specified level range. The RF output level is measured for power level settings of -20 dBm to 10 dBm in 2.5dB step increments at 9 kHz, 100 kHz, 50 MHz, 3 GHz, 4.5 GHz, 5 GHz, and -20 dBm to 9 dBm in 2.5 dB step increments at 6 GHz, and -20 dBm to 8 dBm in 2.5...
  • Page 34: Trace Noise Test

    5. TRACE NOISE TEST Description This test checks the trace noise level for each test port of E5071C. The trace noise level is quantified by performing a “through” measurement 32 times from 9 kHz to 8.5 GHz range, with a cable connected between two ports. The trace noise level for S11, S22, S33 and S44 are quantified by performing a “short”...
  • Page 35: Crosstalk Test

    5. CROSSTALK TEST Description This test checks the crosstalks between test ports of the E5071C. The crosstalk is tested by performing "through" measurements with two test ports connected together and, short-ended "isolation" measurements with the test ports terminated with N-type "Short"...
  • Page 36: System Dynamic Range Test

    6. SYSTEM DYNAMIC RANGE TEST Description This test checks the system dynamic range for the receiver ports of the E5071C. The system dynamic range is tested by performing an "Isolation" measurement 16 times with segment sweep points for specified frequency ranges (after the response and isolation calibrations are performed) and calculating the RMS deviation value from the 16 measurement data for each sweep frequency point.
  • Page 37: Dynamic Accuracy Test

    7. DYNAMIC ACCURACY TEST Description This test checks the dynamic accuracy of the E5071C. The dynamic accuracy is tested at 1.195 GHz for each receiver port using the Agilent Z5623A (with option H01) Dynamic Accuracy Test Kit. The 1.195 GHz source signal of the E5071C goes through the Z5623A's first attenuator variable from 0 dB to 11 dB in 1 dB steps and second attenuator variable up to 100 dB in 10 dB steps and is measured with the receiver port.
  • Page 38: Uncorrected System Performance Test

    Performance Test 8. UNCORRECTED SYSTEM PERFORMANCE TEST 8. UNCORRECTED SYSTEM PERFORMANCE TEST Description This test checks the directivity, source match, load match, transmission tracking, and reflection tracking which are the key S-parameter measurement hardware characteristics. These characteristics are tested using the Agilent 85032F Calibration Kit at 3MHz, 10 MHz, as well as 50 MHz increments from 50 MHz to 3 GHz (up to 8.5 GHz for the E5071B).
  • Page 39: E5071C Performance Test Calculation Sheet

    Performance Test E5071C Performance Test Calculation Sheet E5071C Performance Test Calculation Sheet Introduction This section contains calculation sheets for each performance test that requires additional calculations to determine the final test result. Use the calculation sheet in this section as an aid for recording raw measurement data and calculating the performance test results.
  • Page 40 Performance Test E5071C Performance Test Calculation Sheet Flatness Test ( Option 235, 435 ) Frequency [Hz] Power meter reading [dBm] Test result equation [REF] 50 M —————— 100 k —————— a - REF 300 k —————— a - REF ——————...
  • Page 41 Performance Test E5071C Performance Test Calculation Sheet Frequency [Hz] Power meter reading [dBm] Test result equation [REF] 2.05 G —————— a - REF 2.55 G —————— a - REF 3.00 G —————— a - REF 3.025 G —————— a - REF 3.525 G...
  • Page 42 Performance Test E5071C Performance Test Calculation Sheet (Option 280, 480) Frequency [Hz] Power meter reading [dBm] Test result equation [REF] 50 M —————— —————— a - REF 20 k —————— a - REF 70 k —————— a - REF 100 k ——————...
  • Page 43 Performance Test E5071C Performance Test Calculation Sheet Frequency [Hz] Power meter reading [dBm] Test result equation [REF] 50 M —————— 100 k —————— a - REF 300 k —————— a - REF —————— a - REF 68 M —————— a - REF 68 M + 1 ——————...
  • Page 44: Rf Output Level Linearity Test

    Performance Test E5071C Performance Test Calculation Sheet 4. RF Output Level Linearity Test NOTE Make a copy of the following calculation sheet for each combination of the test frequencies listed below: Option 230, 430: 9 kHz, 100 kHz, 50 MHz, 3 GHz...
  • Page 45 Performance Test E5071C Performance Test Calculation Sheet CW frequency: ______________ Hz Power level Power meter reading [dBm] Test result equation [dBm] —————— —————— a-REF —————— a-REF —————— a-REF —————— a-REF -2.5 —————— a-REF —————— a-REF -7.5 —————— a-REF —————— a-REF -12.5...
  • Page 46 Performance Test E5071C Performance Test Calculation Sheet Power level Power meter reading [dBm] Test result equation [dBm] -12.5 —————— a-REF —————— a-REF -17.5 —————— a-REF —————— a-REF NOTE Make a copy of the following calculation sheet for each combination of the test frequencies listed below: Option 280, 285, 480, 485: 8.5 GHz...
  • Page 47: Dynamic Accuracy Test

    Performance Test E5071C Performance Test Calculation Sheet 8. Dynamic Accuracy Test Make a copy of the following calculation sheet for each of the S-parameters listed below: Option 230, 235, 240, 245, 280 and 285: S21 and S12 Option 430, 435, 440, 445, 480 and 485: S21, S12, S31 and S41 NOTE A dynamic accuracy calculation example is shown at the end of the calculation sheets.
  • Page 48 Performance Test E5071C Performance Test Calculation Sheet Name Power Pm [dB] Tx [dB] @ Port (Port Char Running True setting OFF) D.A. D.A. [dB] DA45 ——— ——— ( Tx @ DA45) - ( Tx @ DA40b) - (Pm @ DA45) + (Pm @ DA40b) →...
  • Page 49 Performance Test E5071C Performance Test Calculation Sheet Name Power Pm [dB] Tx [dB] @ Port (Port Char Running True setting OFF) D.A. D.A. [dB] Adjust source power for (Pm [db] @ DA80a) - ( Tx [dB] @ DA80a) + ( Tx [dB] @ DA80)
  • Page 50: E5071C Performance Test Record

    Performance Test E5071C Performance Test Record E5071C Performance Test Record Agilent Technologies E5071C RF Network Analyzer Serial Number: Option: °C Temperature: Test Date: Humidity: % R.H. Tested by: AUX Input Test 1. Multimeter : 3458A AUX1 Offset error AUX Power...
  • Page 51 Performance Test E5071C Performance Test Record AUX Input Test limit [mV] Test result [mV] Measurement Range[V] uncertainty [mV] ± 110 ± 1.3 ± 110 ± 1.2 - 10 2. Multimeter : 34401A, 34410A or 34411A AUX1 Offset error AUX Power...
  • Page 52: Frequency Accuracy Test

    Performance Test E5071C Performance Test Record Frequency Accuracy Test Option 230, 430 Without Option 1E5 Frequency [Hz] Test limit [Hz] Test result [Hz] Measurement uncertainty [Hz] ± 0.045 ± 0.0004 ± 0.5 ± 0.004 100 k ± 250 ± 1.7 50 M ±...
  • Page 53 Performance Test E5071C Performance Test Record With Option 1E5 Frequency [Hz] Test limit [Hz] Test result [Hz] Measurement uncertainty [Hz] ± 0.1 ± 0.003 100 k ± 50 ± 0.58 50 M ± 134.1 ± 1.6 134.1 M ± 0.548 k ±...
  • Page 54 Performance Test E5071C Performance Test Record Without Option 1E5 Frequency [Hz] Test limit [Hz] Test result [Hz] Measurement uncertainty [Hz] ± 0.5 ± 0.004 100 k ± 250 ± 1.7 50 M ± 670.5 ± 4.6 134.1 M ± 2.74 k ±...
  • Page 55 Performance Test E5071C Performance Test Record With Option 1E5 Frequency [Hz] Test limit [Hz] Test result [Hz] Measurement uncertainty [Hz] ± 0.009 ± 0.0003 ± 0.1 ± 0.003 100 k ± 50 ± 0.58 50 M ± 134.1 ± 1.6 134.1 M...
  • Page 56: Rf Output Level Accuracy And Flatness Test

    Performance Test E5071C Performance Test Record RF Output Level Accuracy and Flatness Test 1. Power Sensor: E9304A Option H18 Level Accuracy Test (@ Port 1) Power level Frequency Test limit Test result Measurement [dBm] [Hz] [dB] [dB] uncertainty [dB] ± 0.65 ±...
  • Page 57 Performance Test E5071C Performance Test Record Frequency [Hz] Test limit [dB] Test result [dB] Measurement uncertainty [dB] ± 1.0 ± 0.23 68 M ± 1.0 ± 0.23 68M + 1 ± 1.0 ± 0.24 550 M ± 1.0 ± 0.24 1.05 G...
  • Page 58 Performance Test E5071C Performance Test Record Option 245, 445 Frequency [Hz] Test limit [dB] Test result [dB] Measurement uncertainty [dB] ± 1.0 ± 0.27 100 k ± 1.0 ± 0.27 300 k ± 1.0 ± 0.23 ± 1.0 ± 0.23 68 M ±...
  • Page 59 Performance Test E5071C Performance Test Record Frequency [Hz] Test limit [dB] Test result [dB] Measurement uncertainty [dB] ± 1.0 ± 0.23 2.05 G ± 1.0 ± 0.23 2.55 G ± 1.0 ± 0.23 ± 1.0 ± 0.24 3.025 G ± 1.0 ±...
  • Page 60: Rf Output Level Accuracy And Flatness Test

    Performance Test E5071C Performance Test Record Frequency [Hz] Test limit [dB] Test result [dB] Measurement uncertainty [dB] ± 1.0 ± 0.24 3.025 G ± 1.0 ± 0.24 3.525 G ± 1.0 ± 0.24 4.025 G ± 1.0 ± 0.24 4.525 G ±...
  • Page 61 Performance Test E5071C Performance Test Record Frequency [Hz] Test limit [dB] Test result [dB] Measurement uncertainty [dB] ± 1.0 ± 0.28 300 k ± 1.0 ± 0.25 ± 1.0 ± 0.25 68 M ± 1.0 ± 0.25 68M + 1 ±...
  • Page 62 Performance Test E5071C Performance Test Record Frequency [Hz] Test limit [dB] Test result [dB] Measurement uncertainty [dB] ± 1.0 ± 0.28 70 k ± 1.0 ± 0.28 100 k ± 1.0 ± 0.28 300 k ± 1.0 ± 0.25 ± 1.0 ±...
  • Page 63 Performance Test E5071C Performance Test Record Frequency [Hz] Test limit [dB] Test result [dB] Measurement uncertainty [dB] ± 1.0 ± 0.25 ± 1.0 ± 0.25 3.025 G ± 1.0 ± 0.25 3.525 G ± 1.0 ± 0.25 4.025 G ± 1.0 ±...
  • Page 64 Performance Test E5071C Performance Test Record Frequency [Hz] Test limit [dB] Test result [dB] Measurement uncertainty [dB] ± 1.0 ± 0.30 6.525 G ± 1.0 ± 0.30 7.025 G ± 1.0 ± 0.30 7.525 G ± 1.0 ± 0.30 8.025 G ±...
  • Page 65: Rf Output Level Linearity Test (@ Port1)

    Performance Test E5071C Performance Test Record Frequency [Hz] Test limit [dB] Test result [dB] Measurement uncertainty [dB] ± 1.0 ± 0.30 8.025 G ± 1.0 ± 0.30 8.5 G RF Output Level Linearity Test (@ Port1) 1. Power Sensor: E9304A Option H18...
  • Page 66 Performance Test E5071C Performance Test Record Power level Test limit [dB] Test result [dB] Measurement [dBm] uncertainty [dB] ± 0.75 ± 0.22 -7.5 ± 0.75 ± 0.22 ± 0.75 ± 0.22 -2.5 ± 0.75 ± 0.21 ± 0.75 ± 0.21 ±...
  • Page 67 Performance Test E5071C Performance Test Record Power level Test limit [dB] Test result [dB] Measurement [dBm] uncertainty [dB] ± 0.75 ± 0.23 ± 0.75 ± 0.22 -7.5 ± 0.75 ± 0.22 ± 0.75 ± 0.22 -2.5 ± 0.75 ± 0.21 ±...
  • Page 68 Performance Test E5071C Performance Test Record Power level Test limit [dB] Test result [dB] Measurement [dBm] uncertainty [dB] ± 0.75 ± 0.22 -17.5 ± 0.75 ± 0.22 ± 0.75 ± 0.22 -12.5 ± 0.75 ± 0.22 ± 0.75 ± 0.22 -7.5...
  • Page 69 Performance Test E5071C Performance Test Record CW Frequency: 6 GHz (relative to 0 dBm reference) Option 280, 285, 480, 485 Power level Test limit [dB] Test result [dB] Measurement [dBm] uncertainty [dB] ± 0.75 ± 0.23 ± 0.75 ± 0.23 -17.5...
  • Page 70: Rf Output Level Linearity Test (@ Port1)

    Performance Test E5071C Performance Test Record Power level Test limit [dB] Test result [dB] Measurement [dBm] uncertainty [dB] ± 0.75 ± 0.21 CW Frequency: 8.5 GHz (relative to 0 dBm reference) Option 280, 285, 480, 485 Power level Test limit [dB]...
  • Page 71 Performance Test E5071C Performance Test Record Power level Test limit [dB] Test result [dB] Measurement [dBm] uncertainty [dB] ± 0.75 ± 0.23 -2.5 ± 0.75 ± 0.22 ± 0.75 ± 0.22 ± 0.75 ± 0.22 ± 0.75 ± 0.22 CW Frequency: 100 kHz (relative to 0 dBm reference)
  • Page 72 Performance Test E5071C Performance Test Record Power level Test limit [dB] Test result [dB] Measurement [dBm] uncertainty [dB] ± 0.75 ± 0.23 ± 0.75 ± 0.23 -2.5 ± 0.75 ± 0.22 ± 0.75 ± 0.22 ± 0.75 ± 0.22 ± 0.75 ±...
  • Page 73 Performance Test E5071C Performance Test Record Power level Test limit [dB] Test result [dB] Measurement [dBm] uncertainty [dB] ± 0.75 ± 0.24 ± 0.75 ± 0.24 -7.5 ± 0.75 ± 0.24 ± 0.75 ± 0.24 -2.5 ± 0.75 ± 0.24 ±...
  • Page 74 Performance Test E5071C Performance Test Record CW Frequency: 6 GHz (relative to 0 dBm reference) Option 280, 285, 480, 485 Power level Test limit [dB] Test result [dB] Measurement [dBm] uncertainty [dB] ± 0.75 ± 0.24 ± 0.75 ± 0.24 -17.5...
  • Page 75: Trace Noise Test

    Performance Test E5071C Performance Test Record Power level Test limit [dB] Test result [dB] Measurement [dBm] uncertainty [dB] ± 0.75 ± 0.23 CW Frequency: 8.5 GHz (relative to 0 dBm reference) Option 280, 285, 480, 485 Power level Test limit [dB]...
  • Page 76 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 10 M - 3 G 70 k < 0.035 0.001 Direction: S22 (Magnitude) Frequency IF bandwidth Test limit...
  • Page 77 Performance Test E5071C Performance Test Record Direction: S12 (Magnitude) Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [dB rms] [dB rms] uncertainty [dB] 9 k - 30 k < 0.004 0.00009 30 k - 10 M < 0.003 0.00009...
  • Page 78 Performance Test E5071C Performance Test Record Direction: S22 (Phase) Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 100 k - 10 M < 0.02 0.0006 10 M - 3 G 70 k <...
  • Page 79 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [dB rms] [dB rms] uncertainty [dB] 30 k - 10 M < 0.003 0.00009 10 M - 4.38 G 70 k < 0.004 0.00033 4.38 G - 4.5 G...
  • Page 80 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [dB rms] [dB rms] uncertainty [dB] 10 M - 4.38 G 70 k < 0.004 0.00033 4.38 G - 4.5 G 70 k < 0.006 0.00023...
  • Page 81 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [dB rms] [dB rms] uncertainty [dB] 4.38 G - 4.5 G 70 k < 0.006 0.00023 Direction: S11 (Phase) Frequency IF bandwidth Test limit...
  • Page 82 Performance Test E5071C Performance Test Record Direction: S21 (Phase) Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 100 k - 10 M < 0.02 0.0006 10 M - 4.38 G 70 k <...
  • Page 83 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 30 k - 10 M < 0.02 0.0006 10 M - 3 G 70 k < 0.035 0.001 3 G - 4.38 G...
  • Page 84 Performance Test E5071C Performance Test Record Direction: S21 (Phase) Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 9 k - 30 k < 0.035 0.00077 30 k - 10 M < 0.02 0.0006...
  • Page 85 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [dB rms] [dB rms] uncertainty [dB] 4.38 G - 8.5 G 70 k < 0.006 0.00023 Direction: S11 (Phase) Frequency IF bandwidth Test limit...
  • Page 86 Performance Test E5071C Performance Test Record Direction: S21 (Phase) Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 100 k - 10 M < 0.02 0.004 10 M - 3 G 70 k <...
  • Page 87 Performance Test E5071C Performance Test Record Direction: S11 (Phase) Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 9 k - 30 k < 0.035 0.00077 30 k - 10 M < 0.02 0.0006...
  • Page 88 Performance Test E5071C Performance Test Record Direction: S44 (Magnitude) Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [dB rms] [dB rms] uncertainty [dB] 9 k - 30 k < 0.004 0.00009 30 k - 10 M < 0.003 0.00009...
  • Page 89 Performance Test E5071C Performance Test Record Direction: S12 (Phase) Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 9 k - 30 k < 0.035 0.00077 30 k - 10 M < 0.02 0.0006...
  • Page 90 Performance Test E5071C Performance Test Record Option 435 Direction: S11 (Magnitude) Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [dB rms] [dB rms] uncertainty [dB] 100 k - 10 M < 0.003 0.00009 10 M - 3 G 70 k <...
  • Page 91 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 10 M - 3 G 70 k < 0.035 0.001 Direction: S44 (Magnitude) Frequency IF bandwidth Test limit...
  • Page 92 Performance Test E5071C Performance Test Record Direction: S12 (Phase) Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 100 k - 10 M < 0.02 0.0006 10 M - 3 G 70 k <...
  • Page 93 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [dB rms] [dB rms] uncertainty [dB] 30 k - 10 M < 0.003 0.00009 10 M - 3 G 70 k < 0.004 0.00033 3 G - 4.38 G...
  • Page 94 Performance Test E5071C Performance Test Record Direction: S33 (Magnitude) Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [dB rms] [dB rms] uncertainty [dB] 9 k - 30 k < 0.004 0.00009 30 k - 10 M < 0.003 0.00009...
  • Page 95 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 4.38 G - 4.5 G 70 k < 0.050 0.0023 Direction: S21 (Magnitude) Frequency IF bandwidth Test limit...
  • Page 96 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 30 k - 10 M < 0.02 0.0006 10 M - 3 G 70 k < 0.035 0.001 3 G - 4.38 G...
  • Page 97 Performance Test E5071C Performance Test Record Direction: S34 (Phase) Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 9 k - 30 k < 0.035 0.00077 30 k - 10 M < 0.02 0.0006...
  • Page 98 Performance Test E5071C Performance Test Record Direction: S22 (Phase) Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 100 k - 10 M < 0.02 0.0006 10 M - 3 G 70 k <...
  • Page 99 Performance Test E5071C Performance Test Record Direction: S44 (Phase) Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 100 k - 10 M < 0.02 0.0006 10 M - 3 G 70 k <...
  • Page 100 Performance Test E5071C Performance Test Record Direction: S12 (Phase) Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 100 k - 10 M < 0.02 0.0006 10 M - 3 G 70 k <...
  • Page 101 Performance Test E5071C Performance Test Record Direction: S34 (Phase) Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 100 k - 10 M < 0.02 0.0006 10 M - 3 G 70 k <...
  • Page 102 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [dB rms] [dB rms] uncertainty [dB] 4.38 G - 8.5 G 70 k < 0.006 0.00023 Direction: S22 (Phase) Frequency IF bandwidth Test limit...
  • Page 103 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [dB rms] [dB rms] uncertainty [dB] 30 k - 10 M < 0.003 0.00009 10 M - 3 G 70 k < 0.004 0.00033 3 G - 4.38 G...
  • Page 104 Performance Test E5071C Performance Test Record Direction: S12 (Magnitude) Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [dB rms] [dB rms] uncertainty [dB] 9 k - 30 k < 0.004 0.00009 30 k - 10 M < 0.003 0.00009...
  • Page 105 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 4.38 G - 8.5 G 70 k < 0.05 0.0023 Direction: S34 (Magnitude) Frequency IF bandwidth Test limit...
  • Page 106 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 10 M - 3 G 70 k < 0.035 0.001 3 G - 4.38 G 70 k < 0.035 0.001...
  • Page 107 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 3 G - 4.38 G 70 k < 0.035 0.001 4.38 G - 8.5 G 70 k < 0.05 0.0023...
  • Page 108 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [deg rms] [deg rms] Uncertaintiy [deg] 4.38 G - 8.5 G 70 k < 0.05 0.0023 Direction: S12 (Magnitude) Frequency IF bandwidth Test limit...
  • Page 109: Crosstalk Test

    Performance Test E5071C Performance Test Record Direction: S34 (Magnitude) Frequency IF bandwidth Test limit Test result Measurement Range[Hz] [Hz] [dB rms] [dB rms] uncertainty [dB] 100 k - 10 M < 0.003 0.00009 10 M - 3 G 70 k <...
  • Page 110 Performance Test E5071C Performance Test Record Option 235 Direction: S21 Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 100 k - 300 k < -100 1.73 300 k - 10 M < -110 3.63...
  • Page 111 Performance Test E5071C Performance Test Record Option 245 Direction: S21 Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 100 k - 300 k < -100 1.73 300 k - 10 M < -110 3.63...
  • Page 112 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 100 k - 300 k < -100 1.73 300 k - 10 M < -110 3.63 10 M - 3 G <...
  • Page 113 Performance Test E5071C Performance Test Record Option 430 Direction: S21 Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 9 k - 100 k < -100 1.73 100 k - 300 k < -100 1.73...
  • Page 114 Performance Test E5071C Performance Test Record Option 435 Direction: S21 Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 100 k - 300 k < -100 1.73 300 k - 10 M < -110 3.63...
  • Page 115 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 300 k - 10 M < -110 3.63 10 M - 3 G < -120 4.95 3 G - 4.5 G <...
  • Page 116 Performance Test E5071C Performance Test Record Option 445 Direction: S21 Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 100 k - 300 k < -100 1.73 300 k - 10 M < -110 3.63...
  • Page 117 Performance Test E5071C Performance Test Record Option 480 Direction: S21 Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 9 k - 100 k < -100 1.73 100 k - 300 k < -100 1.73...
  • Page 118 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 5 G - 6G < -110 3.59 6 G - 7 G < -100 2.61 7 G - 8.5 G <...
  • Page 119 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 300 k - 10 M < -110 3.63 10 M - 3 G < -120 4.95 3 G- 5 G <...
  • Page 120: System Dynamic Range Test

    Performance Test E5071C Performance Test Record System Dynamic Range Test Option 230 Direction S21 Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 9 k - 100 k < -97 1.52 100 k - 300 k <...
  • Page 121 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 100 k - 300 k < -72 1.52 300 k - 10 M < -82 1.13 10 M - 3 G <...
  • Page 122 Performance Test E5071C Performance Test Record Direction S12 Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 9 k - 100 k < -97 1.52 100 k - 300 k < -97 1.52 300 k - 10 M <...
  • Page 123 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 3 G - 4.5 G < -123 1.24 100 k - 300 k < -72 1.52 300 k - 10 M <...
  • Page 124 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 100 k - 300 k < -97 1.52 300 k - 10 M < -107 1.26 10 M - 3 G <...
  • Page 125 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 5 G - 6 G < -98 1.08 6 G - 7 G < -92 1.10 7 G - 8.5 G <...
  • Page 126 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 100 k - 300 k < -72 1.52 300 k - 10 M < -82 1.13 10 M - 3 G <...
  • Page 127 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 100 k - 300 k < -97 1.52 300 k - 10 M < -107 1.26 10 M - 3 G <...
  • Page 128 Performance Test E5071C Performance Test Record Direction S43 Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 100 k - 300 k < -97 1.52 300 k - 10 M < -107 1.26 10 M - 3 G <...
  • Page 129 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 3 G - 4.5 G < -98 1.08 Direction S32 Frequency IF bandwidth Test limit Test result Measurement range [Hz]...
  • Page 130 Performance Test E5071C Performance Test Record Direction S14 Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 9 k - 100 k < -97 1.52 100 k - 300 k < -97 1.52 300 k - 10 M <...
  • Page 131 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 3 G - 4.5 G < -123 1.24 100 k - 300 k < -72 1.52 300 k - 10 M <...
  • Page 132 Performance Test E5071C Performance Test Record Option 480 Direction S21 Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 9 k - 100 k < -97 1.52 100 k - 300 k < -97 1.52...
  • Page 133 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 9 k - 100 k < -72 1.52 100 k - 300 k < -72 1.52 300 k - 10 M <...
  • Page 134 Performance Test E5071C Performance Test Record Direction S14 Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 9 k - 100 k < -97 1.52 100 k - 300 k < -97 1.52 300 k - 10 M <...
  • Page 135 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 10 M - 3 G < -98 1.08 3 G - 5 G < -98 1.08 5 G - 6 G <...
  • Page 136 Performance Test E5071C Performance Test Record Frequency IF bandwidth Test limit Test result Measurement range [Hz] [Hz] [dB] [dB] uncertainty [dB] 5 G - 6 G < -123 1.16 6 G - 7 G < -117 1.08 7 G - 8.5 G <...
  • Page 137: Dynamic Accuracy Test

    Performance Test E5071C Performance Test Record Dynamic Accuracy Test @ Reference power level: -10 dBm Direction: S21 Power level [dBm] Test limit [dB] Test result [dB] Measurement uncertainty [dB] ± 0.207 ± 0.010 ± 0.075 ± 0.010 ± 0.041 ± 0.007 ±...
  • Page 138 Performance Test E5071C Performance Test Record Power level [dBm] Test limit [dB] Test result [dB] Measurement uncertainty [dB] ± 0.041 ± 0.007 ± 0.031 ± 0.009 ± 0.030 ± 0.006 ± 0.035 ± 0.007 ± 0.040 ± 0.010 ± 0.045 ±...
  • Page 139 Performance Test E5071C Performance Test Record Power level [dBm] Test limit [dB] Test result [dB] Measurement uncertainty [dB] ± 0.040 ± 0.010 ± 0.045 ± 0.011 ± 0.056 ± 0.015 ± 0.067 ± 0.019 ± 0.078 ± 0.021 ± 0.091 ±...
  • Page 140: Uncorrected System Performance Test

    Performance Test E5071C Performance Test Record Power level [dBm] Test limit [dB] Test result [dB] Measurement uncertainty [dB] ± 0.078 ± 0.021 ± 0.091 ± 0.021 ± 0.106 ± 0.024 ± 0.125 ± 0.024 ± 0.151 ± 0.026 ± 0.189 ±...
  • Page 141 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] ± 1.0 ± 0.04 Reflection 300 k - 3G Tracking Port 2 System Frequency Test limit Test result Measurement performance...
  • Page 142 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] ± 1.0 ± 0.20 Transmission 300 k - 3 G Tracking ± 1.5 ± 0.04 Reflection 100 k - 300 k Tracking ±...
  • Page 143 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] Source match 3 G - 4.5 G < -20 1.99 Load match 9 k - 300 k < -12 0.04...
  • Page 144 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] ± 1.0 ± 0.24 Transmission 3 G - 4.5 G Tracking ± 1.5 ± 0.03 Reflection 9 k - 300 k Tracking ±...
  • Page 145 Performance Test E5071C Performance Test Record Port 2 System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] Directivity 100 k - 300 k < -20 0.09 Directivity 300 k -3 G < -25 0.81 Directivity 3 G - 4.5 G...
  • Page 146 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] Source match 3 G - 6 G < -20 1.94 Source match 6 G - 8.5 G < -15 0.57...
  • Page 147 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] Load match 9 k - 300 k < -12 0.04 Load match 300 k - 3 G < -17 0.59...
  • Page 148 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] Load match 300 k - 3 G < -17 0.59 Load match 3 G - 6 G < -12 0.66 Load match 6 G - 8.5 G...
  • Page 149 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] Load match 6 G - 8.5 G < -10 0.64 ± 1.5 ± 0.03 Transmission 100 k - 300 k Tracking ±...
  • Page 150 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] ± 1.0 ± 0.04 Reflection 300 k - 3 G Tracking Port 2 System Frequency Test limit Test result Measurement...
  • Page 151 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] ± 1.5 ± 0.04 Reflection 9 k - 300 k Tracking ± 1.0 ± 0.04 Reflection 300 k - 3 G...
  • Page 152 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] ± 1.5 ± 0.03 Transmission 100 k - 300 k Tracking ± 1.0 ± 0.20 Transmission 300 k - 3 G Tracking ±...
  • Page 153 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] Load match 100 k - 300 k < -12 0.04 Load match 300 k - 3 G < -17 0.59 ±...
  • Page 154 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] Directivity 300 k - 3 G < -25 0.81 Directivity 3 G - 4.5 G < -20 0.40 Source match 9 k - 300 k <...
  • Page 155 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] ± 1.5 ± 0.03 Transmission 9 k - 300 k Tracking ± 1.0 ± 0.2 Transmission 300 k - 3 G Tracking ±...
  • Page 156 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] ± 1.0 ± 0.66 Reflection 3 G - 4.5 G Tracking Port 4 System Frequency Test limit Test result Measurement...
  • Page 157 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] Directivity 3 G - 4.5 G < -20 0.40 Source match 100 k - 300 k < -20 0.05 Source match 300 k -3 G <...
  • Page 158 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] ± 1.5 ± 0.03 Transmission 100 k - 300 k Tracking ± 1.0 ± 0.20 Transmission 300 k - 3 G Tracking ±...
  • Page 159 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] ± 1.0 ± 0.66 Reflection 3 G - 4.5 G Tracking Port 4 System Frequency Test limit Test result Measurement...
  • Page 160 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] Directivity 3 G - 6 G < -20 0.40 Directivity 6 G - 8.5 G < -15 0.67 Source match 9 k - 300 k <...
  • Page 161 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] Source match 9 k - 300 k < -20 0.05 Source match 300 k -3 G < -25 3.00 Source match 3 G - 6 G <...
  • Page 162 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] Source match 3 G - 6 G < -20 1.94 Source match 6 G - 8.5 G < -15 0.57...
  • Page 163 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] Load match 9 k - 300 k < -12 0.04 Load match 300 k - 3 G < -17 0.59...
  • Page 164 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] Load match 300 k - 3 G < -17 0.59 Load match 3 G - 6 G < -12 0.66 Load match 6 G - 8.5 G...
  • Page 165 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] Load match 6 G - 8.5 G < -10 0.64 ± 1.5 ± 0.03 Transmission 100 k - 300 k Tracking ±...
  • Page 166 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] ± 1.5 ± 0.03 Transmission 100 k - 300 k Tracking ± 1.0 ± 0.20 Transmission 300 k - 3 G Tracking ±...
  • Page 167 Performance Test E5071C Performance Test Record System Frequency Test limit Test result Measurement performance range [Hz] [dB] [dB] uncertainty [dB] ± 1.0 ± 0.20 Transmission 300 k - 3 G Tracking ± 1.0 ± 0.24 Transmission 3 G - 6 G Tracking ±...
  • Page 168 Performance Test E5071C Performance Test Record Chapter 2...
  • Page 169: Adjustment

    This chapter provides the adjustment information for the E5071C ENA Series Network Analyzer to ensure that it is within its specifications. The adjustment must be performed Agilent’s qualified service personnel. If you need the adjustment for your E5071C, it should be sent to the nearest Agilent Technologies service office.
  • Page 170: Safety Considerations

    Intentional interruption of the protective ground system for any reason is prohibited. Warm-up for Adjustment Warm-up the E5071C for at least 30 minute before performing any of the following adjustment procedures to ensure proper results and correct instrument operation. Required Equipment Table 1-1 on page 20 lists the equipment required to perform the adjustment procedures described in this chapter.
  • Page 171: Required Adjustment After Replacing Assembly

    After replacing the following assembly, the adjustment items described in Table 3-1 must be required. The adjustment must be performed Agilent’s qualified service personnel. If you need the adjustment for your E5071C, it should be sent to the nearest Agilent Technologies service office.
  • Page 172: Frequency Reference Adjustment

    Adjustment Required Adjustment after Replacing Assembly Required equipment for frequency reference adjustment Description Recommended Model Frequency Counter Agilent 53181A with Opt. 010 and 124 or Agilent 53131/2A with Opt. 010 and 124 Frequency Standard Agilent E5071A BNC(m)-BNC(m) Cable, 61 cm Agilent p/n 8120-1839 Frequency Reference Adjustment The purpose of this procedure is to adjust the 10 MHz frequency reference on the source...
  • Page 173: Virtual Bridge Coefficient Adjustment

    Adjustment Required Adjustment after Replacing Assembly Virtual Bridge Coefficient Adjustment This item writes the coefficient factors to the E5071C. Required equipment for the source power adjustment None Source Output Power Adjustment The purpose of this procedure is to adjust the variable attenuator in Level Vernier board, and the gains among Level Vernier, Distributor and Receiver.
  • Page 174: Receiver Rf Range Adjustment

    Adjustment Required Adjustment after Replacing Assembly Receiver RF Range Adjustment This item writes the coefficient of RF Range. Required equipment for the test port characteristics adjustment None Chapter 3...
  • Page 175: Troubleshooting

    Troubleshooting This chapter provides procedure to isolate a faulty assembly in the E5071C Network Analyzer.
  • Page 176: Introduction

    Troubleshooting Introduction Introduction WARNING These servicing instructions are for use by qualified personnel only. To avoid possible electrical shock, do not perform any servicing unless you are qualified to do so. WARNING The opening of covers or removal of parts is likely to expose dangerous voltages. Disconnect the instrument from its power supply beforehand.
  • Page 177: How To Exit From The E5071C Measurement View

    If you wish to return to the Measurement View, double-click "Network Analyzer" icon. NOTE If you need to shut down the E5071C and again turn on, perform in accordance with the following procedure. a. To get "Start" menu bar displayed, move the pointer to the bottom of the screen with mouse.
  • Page 178: To Troubleshoot The Instrument

    185. Step 2. Check the display • If no display appears on the LCD after the E5071C is turned on, go to “No Display troubleshooting” on page 180. • If the E5071C stops in booting process despite something being displayed on the LCD, go to “Boot Process Troubleshooting”...
  • Page 179 Troubleshooting To Troubleshoot the Instrument Figure 4-2 Primary trouble isolation flowchart Display and Boot-up Problems Problems on Front Panel, Keyboard, Mouse, LCD Display, Hard Disk Drive, Video Output, GPIB Interface Power on and External Trigger Input. Go to No Display Display on screen? Troubleshooting Power on...
  • Page 180: No Display Troubleshooting

    No Display troubleshooting No Display troubleshooting If the E5071C displays nothing despite it is powered from proper ac power line, isolate the failure in accordance with the procedure shown in Figure 4-3. Connect the keyboard to the E5071C rear panel USB connector, turn the power on and start trouble isolation.
  • Page 181 Step 2. Check system fans inside. If the system fans on the chassis inside the E5071C don't run, problem seems in the analog base module. In this case, remove the E5071C outer cover and make sure whether the fans run or not.
  • Page 182: Boot Process Troubleshooting

    Troubleshooting Boot Process Troubleshooting Boot Process Troubleshooting Figure 4-4 represents the booting process flow in the E5071C. If the E5071C stops in the booting process, troubleshoot using the following step-by-step procedure. Figure 4-4 Booting process flowchart Power On VIDEO BIOS Title Screen...
  • Page 183 Troubleshooting Boot Process Troubleshooting Step 1. Splash Screen The splash screen is displayed with Agilent logo as shown in Figure 4-5. If the splash screen is displayed, you can assume that the CPU Mother Board Assembly is functioning correctly. NOTE While the splash screen is displayed, if you want to run the BIOS setup utility, push F2 key as soon as in the screen.
  • Page 184 The operating system automatically checks the device drivers, which are necessary to use the E5071C functions and are installed in the system before the E5071C is shipped from Agilent factory. If the operating system doesn't detect them, a message box is displayed. In this case, install the device driver.
  • Page 185 Troubleshooting Boot Process Troubleshooting NOTE If a message of "Will Shut Down in Five Seconds" is displayed in place of "Initializing.." and the shutdown occurs, the PCA PCI DSP card fails in starting up. The following message may be displayed before the shutdown occurs: "Fatal Error: Failed to Initialize DSP Driver": or "Fatal Error: Failed to Initialize DSP": This message indicates that the PCA PCI DSP Card doesn't work or is not properly...
  • Page 186: Troubleshooting Using Diagnostics Test

    Power On Self Test Power-on self-test always takes place once after the E5071C is turned on. When a failure is detected, a message of the "Power on self test failed" is displayed. The content of the power-on self-test is the same as a portion of the siagnostics test program and includes the dc power supply voltage, source PLL synthesizer and level controller (ALC) tests.
  • Page 187: Required Test Equipment

    Troubleshooting Troubleshooting Using Diagnostics Test Table 4-1 Diagnostic test group menu Number Test Name INTERNAL LEVEL MONITOR TEST RF OUTPUT LEVEL POWER SEEP RF OUTPUT LEVEL RANGE TEST RECEIVER ABSOLUTE MEASUREMENT TEST RECEIVER COMPRESSION TEST RECEIVER IF RANGING TEST Required Test Equipment Table 4-2 shows the equipment required for performing the diagnostic test.
  • Page 188 Troubleshooting Troubleshooting Using Diagnostics Test Figure 4-9 Open dialog box e5071cse1062 Step 4. Select "User [D]" (preset state) from menu in the "Lock in:" box. Step 5. Double-click "Agilent" folder to open it and to access its menu. Step 6. Double-click "Service" folder to open it. Step 7.
  • Page 189 Troubleshooting Troubleshooting Using Diagnostics Test Step 12. When the diagnostics test is completed, the dialog box displays Pass or Fail for each test group as an example shown in Figure 4-11. Figure 4-11 Diagnostics test result display example (Pass) Step 13. To exit the siagnostics test, click "Exit" button. NOTE Test result is stocked as “resultDT.txt”...
  • Page 190: Diagnostics Test Failure Troubleshooting

    Troubleshooting Troubleshooting Using Diagnostics Test Diagnostics Test Failure Troubleshooting Table 4-3 represents the contents of the diagnostics tests and the relationships of failed tests to probable faulty board assemblies. If the instrument fails the diagnostics test, replace the faulty board assembly as shown in Table 4-3. It is recommended to perform the diagnostic test in accordance with the main menu.
  • Page 191 Troubleshooting Troubleshooting Using Diagnostics Test NOTE The meaning of abbreviations are shown in Table 4-4. Table 4-4 The meaning of abbreviations Abbreviation Meaning (Description) Analog Base Board TESTED Synthesizer Module (Source) TESTED Synthesizer Module (Local) TESTED Level Vervier Module TESTED Receiver Module NOTE Module configulation is shown Figure 4-13 for 2-port, and Figure 4-14 for 4-port.
  • Page 192 Troubleshooting Troubleshooting Using Diagnostics Test Figure 4-14 Module configulation for 4-port Slot 2: Synthesizer Module for Source Slot 3: Synthesizer Module for Local Slot 6: Receiver Module for Port 1 Slot 7: Receiver Module for Port 2 Slot 8: Receiver Module for Port 3 Slot 9: Receiver Module for Port 4 Chapter 4...
  • Page 193 Troubleshooting Troubleshooting Using Diagnostics Test Figure 4-15 Flowchart for troubleshooting on DC-BUS Test DC-BUS TEST Sequential Error ALL 0V? No Rec B'd? Add Receiver B'd Replace ALL xxV? Analog Base B'd Replace BAS:GND FAIL? Analog Base B'd 8< room Replace BAS:TEMP FAIL? temperature <...
  • Page 194 Troubleshooting Troubleshooting Using Diagnostics Test Figure 4-16 Flowchart for troubleshooting on Internal Level Monitor Test Internal Level Monitor Test Assy 50Ohm on LCL SYN:LCL: FAIL? B'd Output Replace SYN:LCL FAIL? LCL B'd Cable Assy or Distributor B'd has trouble Assy 50Ohm on SRC SYN:SRC: FAIL? B'd Output Port Replace...
  • Page 195: Function Specific Troubleshooting

    Function Specific Troubleshooting Function Specific Troubleshooting If the E5071C exhibits a failure symptom that is related to a specific function or control such as a front panel key control, display, data storage, remote control interface, external trigger, external keyboard or mouse, isolate the trouble using the Function Specific Troubleshooting procedures described below.
  • Page 196: To Check The Device Driver

    Function Specific Troubleshooting To Check the Device Driver Make sure first whether the E5071C device drivers are installed properly or not by the following procedure, if a function of specific device in the E5071C doesn't work. Step 1. Exit from the E5071C measurement view in accordance with the procedure described in “How to exit from the E5071C Measurement View”...
  • Page 197 Troubleshooting Function Specific Troubleshooting necessary device drivers and displays the device names as shown in Figure 4-19. Figure 4-19 System Properties Window (Hardware) e5071cse1065 Click the icon with the right button and click Property to show the detail of the status. as shown in Figure 4-20.
  • Page 198: To Check The Front Panel

    Troubleshooting Function Specific Troubleshooting To Check the Front Panel Procedure Randomly press the front panel keys and rotate the knob to verify that they work normally. Step 1. Press key. Step 2. Click and, then, in the softkeys. Service Menu Test Menu Step 3.
  • Page 199: To Check The Lcd

    White and Black every 2 seconds and returns to the measurement view. If the color test screen doesn't appear correctly, perform step 4. Step 4. Connect an external VGA monitor to the VIDEO output port on the E5071C rear panel. •...
  • Page 200: To Check The Video Output

    To Check the Video Output Procedure Step 1. Connect an external XGA color monitor to the Video output port on the E5071C rear panel. Step 2. Turn the external monitor on. Step 3. Verify that the monitor screen view is the same as the display on the LCD. If the monitor screen view is abnormal, a failure seems in the PCA PCI DSP Card.
  • Page 201: To Check The Removable Hard Disk Drive

    Troubleshooting Function Specific Troubleshooting To Check the Removable Hard Disk Drive Procedure When the power is turned on, it is required to confirm the notch geometry is set as follows. NOTE If the notch geometry is not set as follows, the error message “OS not found” will be shown when you turn on the power.
  • Page 202: Performance Test Failure Troubleshooting

    Performance test failure troubleshooting Performance test failure troubleshooting This section describes the adjustment and troubleshooting procedures when the E5071C fails the performance tests. If the performance of the instrument is critical for the test limits and seems to be adjustable, perform first the adjustment(s) related to the failed test. When the test result is far from the tolerance of the test or the performance is not adjustable, isolate the faulty assembly in accordance with the "Performance tests failure...
  • Page 203: Performance Test Failure Troubleshooting

    Troubleshooting Performance test failure troubleshooting ###: Most suspicious assembly ##: Suspicious assembly #: Possible faulty assembly Performance test failure troubleshooting Table 4-7 represents the relationships between the failed test and probable faulty assembly. If the performance test failure cannot be removed by a proper adjustment, replace the assembly shown in this table.
  • Page 204 Troubleshooting Performance test failure troubleshooting Table 4-7 Performance test failure troubleshooting information First failed test Probable faulty board assembly or parts OCXO -SRC -LCL System Dynamic Range (S32) System Dynamic Range (S43) System Dynamic Range (S14, S12) Dynamic Accuracy (T1) Dynamic Accuracy (T2) Dynamic Accuracy (T3) Dynamic Accuracy (T4)
  • Page 205: Probable Faulty Board Assembly Or Parts On Adjustment Failure

    Troubleshooting Performance test failure troubleshooting NOTE The meaning of abbreviations are shown in Table 4-8. Table 4-8 The meaning of abbreviations Abbreviation Meaning (Description) OCXO OCXO Board Analog Base Module SYN -SRC TESTED Synthesizer Module for Source SYN -LCL TESTED Synthesizer Module for Local TESTED Level Vervier Module RCV -1 TESTED Receiver Module Port 1...
  • Page 206 Troubleshooting Performance test failure troubleshooting Table 4-9 Adjustment failure troubleshooting information Probable faulty board assembly or parts Failed Failed test item Adjustment Synthesizer vSrcDblDet / vSrcAlcDet Gain vLrcDblDet / vLrcAlcDet Source Output ccSrcVnr{1|2}Vref Power RchGain ccSrcATT{1|2| matPowSrc[1000] matPowSrc[4000] Receiver IF ccIFRange[{R1|T1}] Range ccIFRange[{21|T2}]...
  • Page 207 Troubleshooting Performance test failure troubleshooting NOTE The meaning of abbreviations are shown in Table 4-10. Table 4-10 The meaning of abbreviations Abbreviation Meaning (Description) OCXO OCXO Board Analog Base Module SYN -SRC TESTED Synthesizer Module for Source SYN -LCL TESTED Synthesizer Module for Local TESTED Level Vervier Module RCV -1 TESTED Receiver Module Port 1...
  • Page 208: To Configure The Cpu Mother Board Assembly And Bios

    DRAM count in accordance with the following procedure. Step 1. Connect the external keyboard and mouse to the E5071C rear panel connectors. Step 2. Turn the instrument on. Wait until the E5071C boots up. Do not press any key until the E5071C measurement view appears.
  • Page 209: To Confirm Or Set The Bios Options

    If it is not correct, the CPU Mother Board Assembly is faulty. Step 8. Click "OK" button to close the window. Step 9. Turn the E5071C off before proceeding to the next step “To Confirm or Set the BIOS Options”.
  • Page 210 Troubleshooting To configure the CPU Mother Board Assembly and BIOS Step 6. The BIOS setup utility has primary menus, most of which have sub-menus. Brief descriptions for these menus are provided in Table 4-11. It is recommended not to change any value except for the case of replaceing CPU Mother board assembly..
  • Page 211 Setup Defaults into the BIOS ROM. As a result, almost all of the BIOS options are automatically configured to the optimal settings for the E5071C. Exit Without Saving If you cancel the changes in the BIOS option settings, select "Exit Discarding Changes"...
  • Page 212 Troubleshooting To configure the CPU Mother Board Assembly and BIOS Chapter 4...
  • Page 213: Replaceable Parts

    Replaceable Parts This chapter contains information for ordering replacement parts for the E5071C ENA Series RF Network Analyzers.
  • Page 214: Ordering Information

    Replaceable Parts Ordering Information Ordering Information To order part listed in the replaceable part lists, quote the Agilent part number (with a check digit), indicate the quantity required, and address the order to the nearest Agilent office. The check digit will ensure accurate and timely processing of the order. To order a part not listed in the replaceable part table, include the instrument model number, the description and function of the part, and the quantity of parts required.
  • Page 215: Exchange Assemblies

    Replaceable Parts Exchange Assemblies Exchange Assemblies Under the rebuilt-exchange assembly program, certain factory-repaired and tested assemblies are available on a trade-in basis. These assemblies are offered al lower cost than a new assembly, but meet all factory specifications required of a new assembly. The defective assembly must be returned for credit under the terms of the rebuilt-exchange assembly program.
  • Page 216: Replaceable Parts List

    Replaceable Parts Replaceable Parts List Replaceable Parts List Top View (Major Assemblies) Figure 5-1 Top View (Major Assemblies) e5071cse1002 Chapter 5...
  • Page 217 Replaceable Parts Replaceable Parts List Table 5-1 Top View (Major Assemblies) Ref. Agilent Part Number Qty. Description Desig. See Table 5-23 POWER SUPPLY ASSEMBLY E5070-66525 HANDLER I/O BOARD See Table 5-20 PCA DIGITAL BOARD See Table 5-19 GPIB CARD ASSEMBLY See Table 5-19 PCA PCI DSP CARD ASSEMBLY See Table 5-19...
  • Page 218: Top View (Cables)

    Replaceable Parts Replaceable Parts List Top View (Cables) Figure 5-2 Top View (Cables) e5071cse1003 Chapter 5...
  • Page 219 Replaceable Parts Replaceable Parts List Table 5-2 Top View (Cables) Ref. Agilent Part Number Qty. Description Desig. E5071-61603 FLAT CABLE ASSY E5071-61601 FLAT CABLE ASSY E5071-61609 CABLE ASSY E5071-61605 FLAT CABLE ASSY E5071-61612 FLAT CABLE ASSY E5071-61620 FLAT CABLE ASSY E5071-61604 FLAT CABLE ASSY E5071-61614...
  • Page 220: Top View (Miscellaneous Parts)

    Replaceable Parts Replaceable Parts List Top View (Miscellaneous Parts) Figure 5-3 Top View (Miscellaneous Parts) e5071cse1004 Chapter 5...
  • Page 221 Replaceable Parts Replaceable Parts List Table 5-3 Top View (Miscellaneous Parts) Ref. Agilent Part Number Qty. Description Desig. E5071-00611 PLATE E5071-01205 GUIDE BAR 0515-1382 SCREW-MACH M3.5 x L6 FL T15 5041-9176 TRIM STRIP E5071-01201 BRACKET 0515-1946 SCREW-MACH M3.0 x L6 FL T10 E5071-01212 GUARD 0515-1946...
  • Page 222: Top View (Under Power Supply)

    Replaceable Parts Replaceable Parts List Top View (Under Power Supply) Figure 5-4 Top View (Under Power Supply) e5071cse1005 Table 5-4 Top View (Under Power Supply) Ref. Agilent Part Number Qty. Description Desig. E5071-61606 WIRE ASSY 0515-1946 SCREW-MACH M3.0 x L6 FL T10 E5071-00600 BLACKET 1400-0866...
  • Page 223: Top View (Motherboard And Other Parts)

    Replaceable Parts Replaceable Parts List Top View (Motherboard and Other Parts) Figure 5-5 Top View (Motherboard and Other Parts) e5071cse1006 Chapter 5...
  • Page 224 Replaceable Parts Replaceable Parts List Table 5-5 Top View (Motherboard and Other Parts) Ref. Agilent Part Number Qty. Description Desig. 1420-0356 BATTERY LITHIUM (CR2032), 3V, .22A-HR 0515-0430 SCREW-MACH M3.0 x L6 PN T10 0363-0170 GASKET 10 UNITS 0515-0664 SCREW-MACH M3.0 x L12 PN T10 0515-0430 SCREW-MACH M3.0 x L6 PN T10 E5071-61080...
  • Page 225: Bottom View

    Replaceable Parts Replaceable Parts List Bottom View Figure 5-6 Bottom View e5071cse1087 Chapter 5...
  • Page 226 Replaceable Parts Replaceable Parts List Table 5-6 Bottom View Ref. Agilent Part Number Qty. Description Desig. 0515-1382 SCREW-MACH M3.5 x L6 FL T15 0515-1946 SCREW-MACH M3.0 x L6 FL T10 0515-1946 SCREW-MACH M3.0 x L6 FL T10 (Opt. 430, 435,480,485) 0515-0430 SCREW-MACH M3.0 x L6 PN T10 Chapter 5...
  • Page 227: Front View (Analog)

    Replaceable Parts Replaceable Parts List Front View (Analog) Figure 5-7 Front View (Analog) e5071cse1007 Table 5-7 Front View (Analog) Ref. Agilent Part Number Qty. Description Desig. 5087-7272 SWITCH RF SPDT E5071-61084 DISTRIBUTOR MODULE E5071-62083 LEVEL VERNIER BOARD E5071-69083 LEVEL VERNIER BOARD (Exchange) E5071-62081 SYNTHESIZER BOARD E5071-69081...
  • Page 228 Replaceable Parts Replaceable Parts List Table 5-7 Front View (Analog) Ref. Agilent Part Number Qty. Description Desig. E5071-62082 RECEIVER BOARD (Opt. 430, 480) E5071-69082 RECEIVER BOARD (Exchange Opt. 430, 480) E5071-62092 RECEIVER BOARD (Opt. 435, 485) E5071-69092 RECEIVER BOARD (Exchange Opt. 435, 485) Chapter 5...
  • Page 229: Front View (Semirigid Cables) (Opt. 2Xx)

    Replaceable Parts Replaceable Parts List Front View (Semirigid Cables) (Opt. 2xx) Figure 5-8 Front View (Semirigid Cables) (Opt. 2xx) e5071cse1009 Table 5-8 Front View (Semirigid Cables) (Opt. 2xx) Ref. Agilent Part Number Qty. Description Desig. E5071-61641 RF CABLE ASSY SRGD E5071-61640 RF CABLE ASSY SRGD 0955-0301...
  • Page 230: Front View (Semirigid Cables) (Opt. 4Xx)

    Replaceable Parts Replaceable Parts List Front View (Semirigid Cables) (Opt. 4xx) Figure 5-9 Front View (Semirigid Cables) (Opt. 4xx) e5071cse1011 Table 5-9 Front View (Semirigid Cables) (Opt. 4xx) Ref. Agilent Part Number Qty. Description Desig. E5071-61638 RF CABLE ASSY SRGD E5071-61637 RF CABLE ASSY SRGD E5071-61636...
  • Page 231: Front Panel

    Replaceable Parts Replaceable Parts List Front Panel Figure 5-10 Front Panel 1 14 e5071cse1012 Chapter 5...
  • Page 232 Replaceable Parts Replaceable Parts List Table 5-10 Front Panel Ref. Agilent Part Number Qty. Description Desig. E5071-00222 PANEL SUB E5070-66552 FRONT PANEL KEYBOARD E5071-25101 KEYPAD RUBBER See Table 5-30 STANDBY SWITCH ASSEMBLY 0371-3953 KEY CAP See Table 5-30 PCA PROBE POWER See Table 5-30 PCA FRONT USB E5071-00252...
  • Page 233: Rear View (1)

    Replaceable Parts Replaceable Parts List Rear View (1) Figure 5-11 Rear View (1) e5071cse1013 Table 5-11 Rear View (1) Ref. Agilent Part Number Qty. Description Desig. SCREW-MACH M3.0 x L8 PN T10 0515-0372 I/O SHIELD 8160-1677 0515-0372 SCREW-MACH M3.0 x L8 PN T10 2950-0054 2190-0054 WASHER-LK INTL T...
  • Page 234: Rear View (2)

    Replaceable Parts Replaceable Parts List Rear View (2) Figure 5-12 Rear View (2) e5071cse1092 Table 5-12 Rear View (2) Ref. Agilent Part Number Qty. Description Desig. SCREW-MACH M3.0 x L6 FL T10 0515-1946 0515-0430 SCREW-MACH M3.0 x L6 PN T10 SCREW-MACH M3.0 x L6 FL T10 0515-1946 2950-0035...
  • Page 235: Cover Assembly

    Replaceable Parts Replaceable Parts List Cover Assembly Figure 5-13 Cover Assembly e5071cse1014 Table 5-13 Cover Assembly Ref. Agilent Part Number Qty. Description Desig. E5071-04002 COVER ASSY 5041-9186 STRAP HANDLE FRT 0515-2049 SCREW-MACH M5.0 x L16 FL T20 5041-9173 SIDE TRIM 221.5 5041-9167 FOOT FL 1460-1345...
  • Page 236: Left Side View

    Replaceable Parts Replaceable Parts List Left Side View Figure 5-14 Left Side View e5071cse1083 Table 5-14 Left Side View Ref. Agilent Part Number Qty. Description Desig. E5071-61603 FLAT CABLE ASSY E5071-00607 PLATE 0515-1946 SCREW-MACH M3.0 x L6 FL T10 0515-0430 SCREW-MACH M3.0 x L6 PN T10 0515-2146 SCREW-MACH M3.0 x L4 FL T10...
  • Page 237: Right Side View

    Replaceable Parts Replaceable Parts List Right Side View Figure 5-15 Right Side View e5071cse1084 Table 5-15 Right Side View Ref. Agilent Part Number Qty. Description Desig. E5071-61604 FLAT CABLE ASSY 1400-0611 CLAMP CABLE 0515-0430 SCREW-MACH M3.0 x L6 PN T10 0515-2146 SCREW-MACH M3.0 x L4 FL T10 0515-0430...
  • Page 238: Removable Hard Disk Drive Assembly

    Replaceable Parts Replaceable Parts List Removable Hard Disk Drive Assembly Figure 5-16 Removable Hard Disk Drive Assembly e5071cse1016 Table 5-16 Removable Hard Disk Drive Assembly Ref. Agilent Part Number Qty. Description Desig. 5188-4410 HDD CARRIER (Service Center Only) E5071-82301 HDD INSTALLED (Service Center Only) 0515-4353 SCREW-MACH M3.0 x L3 PN T10 (Service Center Only) 0960-2562...
  • Page 239: Removable Hard Disk Drive Assembly Option 017

    Replaceable Parts Replaceable Parts List Removable Hard Disk Drive Assembly Option 017 Figure 5-17 Removable Hard Disk Drive Assembly Option 017 e5071cse3001 Chapter 5...
  • Page 240 Replaceable Parts Replaceable Parts List Table 5-17 Removable Hard Disk Drive Assembly Ref. Agilent Part Number Qty. Description Desig. E5052-60101 REMOVABLE HDD 2.5INCH RECEIVING FRAME ASSY E5070-66538 CONNECTOR BD PCA E5070-61668 FLAT CBL ASSY E5071-82301 HDD INSTALLED (Service Center Only) E5070-00650 PLATE BOTTOM 0515-2146...
  • Page 241: Hard Disk Drive Assembly Option 019

    Replaceable Parts Replaceable Parts List Hard Disk Drive Assembly Option 019 Figure 5-18 Hard Disk Drive Assembly Option 019 HDD BOTTOM e5071cse3002 Chapter 5...
  • Page 242 Replaceable Parts Replaceable Parts List Table 5-18 Hard Disk Drive Assembly Ref. Agilent Part Number Qty. Description Desig. E5071-82301 HDD INSTALLED (Service Center Only) E5070-66536 CONNECTOR BD PCA E5071-01222 HDD BRACKET 0515-0430 SCREW-MACH M3.0 x L6 PN T10 0515-0374 SCREW-MACH M3.0 x L10 PN T10 0515-2028 SCREW-MACH M2.5 x L6 FL T8 E5070-01255...
  • Page 243: Pci Dsp Card, Gpib Card And Usb Card Assembly

    Replaceable Parts Replaceable Parts List PCI DSP Card, GPIB Card and USB Card Assembly Figure 5-19 PCI DSP Card, GPIB Card AND USB Card Assembly e5071cse1021 Table 5-19 PCI DSP Card, GPIB Card AND USB Card Assembly Ref. Agilent Part Number Qty.
  • Page 244: Pca Digital Board Assembly

    Replaceable Parts Replaceable Parts List PCA Digital Board Assembly Figure 5-20 PCA Digital Board Assembly e5071cse1088 Table 5-20 PCA Digital Board Assembly Ref. Agilent Part Number Qty. Description Desig. E5070-66553 PCA DIGITAL BRIDGE 0515-0430 SCREW-MACH M3.0 x 6 PN T10 E5071-01210 STAY Chapter 5...
  • Page 245: T2 Switch Assembly

    Replaceable Parts Replaceable Parts List T2 Switch Assembly Figure 5-21 T2 Switch Assembly 2-port Option 4-port Option e5071cse1086 Table 5-21 T2 Switch Assembly Ref. Agilent Part Number Qty. Description Desig. E5071-61084 DISTRIBUTOR MODULE E5071-61619 WIRE ASSY (Opt.2xx) WIRE ASSY (Opt.4xx) 5087-7272 SWITCH RF SPDT (Opt.2xx) SWITCH RF SPDT (Opt.4xx)
  • Page 246: Fan Assembly

    Replaceable Parts Replaceable Parts List Fan Assembly Figure 5-22 Fan Assembly e5071cse1089 Table 5-22 Fan Assembly Ref. Agilent Part Number Qty. Description Desig. E5071-01200 BRACKET FAN 0361-1885 RIVET PLASTIC E5071-61618 FAN ASSY 1400-0249 CABLE TIE Chapter 5...
  • Page 247: Power Suplly Assembly

    Replaceable Parts Replaceable Parts List Power Suplly Assembly Figure 5-23 Power Suplly Assembly e5071cse1159 Table 5-23 Power Suplly Assembly Ref. Agilent Part Number Qty. Description Desig. 0950-4773 POWER SUPPLY E5071-00601 PLATE 0515-1946 SCREW-MACH M3.0 x L6 FL T10 Chapter 5...
  • Page 248: Pca Dc Bias And Oven Assembly

    Replaceable Parts Replaceable Parts List PCA DC Bias and Oven Assembly Figure 5-24 PCA DC Bias and Oven Assembly e5071cse1090 Table 5-24 PCA DC Bias and Oven Assembly Ref. Agilent Part Number Qty. Description Desig. E5071-00200 PANEL REAR E5070-66557 PCA DC BIAS 2-PORT (Opt.235, 285) E5070-66558 PCA DC BIAS 4-PORT (Opt.435, 485) 1400-1391...
  • Page 249: Analog Motherboard Assembly

    Replaceable Parts Replaceable Parts List Analog Motherboard Assembly Figure 5-25 Analog Motherboard Assembly e5071cse1085 Table 5-25 Analog Motherboard Assembly Ref. Agilent Part Number Qty. Description Desig. E5070-66550 ANALOG MOTHER BOARD SCREW-MACH M3.0 x 6 PN T10 0515-0430 CLAMP CABLE 1400-1334 E5071-61613 WIRE ASSY 0340-1259...
  • Page 250: Chassis Assembly

    Replaceable Parts Replaceable Parts List Chassis Assembly Figure 5-26 Chassis Assembly e5071cse1024 Chapter 5...
  • Page 251 Replaceable Parts Replaceable Parts List Table 5-26 Chassis Assembly Ref. Agilent Part Number Qty. Description Desig. E5071-60001 CHASSIS 5022-1190 FRONT FRAME 8160-0641 GASKET (125 cm) 0515-2113 SCREW-MACH M4.0 x L8 PN T20 E5071-24007 SPACER 1520-0685 DAMPER 1400-1334 CLAMP CABLE Chapter 5...
  • Page 252: Keyboard Assembly

    Replaceable Parts Replaceable Parts List Keyboard Assembly Figure 5-27 Keyboard Assembly e5071cse1091 Table 5-27 Keyboard Assembly Ref. Agilent Part Number Qty. Description Desig. 0515-0430 SCREW-MACH M3.0 x 6 PN T10 E5070-66552 PCA FRONT KEY 0363-0170 GASKET 19-UNITS Chapter 5...
  • Page 253: Lcd Assembly

    Replaceable Parts Replaceable Parts List LCD Assembly Figure 5-28 LCD Assembly e5071cse1026 Table 5-28 LCD Assembly Ref. Agilent Part Number Qty. Description Desig. 0515-1402 SCREW-MACH M3.5 x L8 PN T15 1400-1391 CLANP CABLE E5070-61628 WIRE ASSY E5070-60112 LCD-TPANEL ASSY See Table 5-29 INVERTER ASSY E5071-61611 WIRE ASSY...
  • Page 254: Inverter Assembly

    Replaceable Parts Replaceable Parts List Inverter Assembly Figure 5-29 Inverter Assembly e5071cse1093 Table 5-29 Inverter Assembly Ref. Agilent Part Number Qty. Description Desig. E5071-01211 BRACKET 0950-4420 INVERTER 0515-1974 SCREW-MACH M2.5 x L4 PN T8 Chapter 5...
  • Page 255: Standby Switch Assembly

    Replaceable Parts Replaceable Parts List Standby Switch Assembly Figure 5-30 Standby Switch Assembly e5071cse1027 Table 5-30 Standby Switch Assembly Ref. Agilent Part Number Qty. Description Desig. 0515-0430 SCREW-MACH M3.0 x L6 PN T10 E5070-66556 PCA PROBE POWER E5070-66555 PCA FRONT USB E5070-61901 WIRE ASSY W/STANDBY SWITCH 0515-2028...
  • Page 256: Cover Assembly (1)

    Replaceable Parts Replaceable Parts List Cover Assembly (1) Figure 5-31 Cover Assembly e5071cse1028 Table 5-31 Cover Assembly Ref. Agilent Part Number Qty. Description Desig. 0515-0430 SCREW-MACH M3.0 x L6 PN T10 1400-3156 CLAMP-CABLE NYLON 6/6 5183-4184 CNTR TOUCH PANEL E5071-00621 COVER 0363-0170 GASKET 10-UNITS...
  • Page 257: Label On Front Face

    Replaceable Parts Replaceable Parts List Label on Front Face Figure 5-32 Label on Front Face ELECTROSTATIC DISCHARGE OR TRANSIENT INPUT VOLTAGE MAY DAMAGE CIRCUITS. e5071cse1160 Table 5-32 Label on Front Face Ref. Agilent Part Number Qty. Description Desig. E5070-87125 LABEL (CAUTION) 1401-0247 CAP-PROT (Opt.
  • Page 258: Label On Rear Face

    Replaceable Parts Replaceable Parts List Label on Rear Face Figure 5-33 Label on Rear Face e5071cse1030 Table 5-33 Label on Rear Face Ref. Agilent Part Number Qty. Description Desig. E5070-87103 LABEL (NOTE for Switch) E5071-87107 LABEL (Opt. 235, 285) E5071-87109 LABEL (Opt.
  • Page 259 Replaceable Parts Replaceable Parts List Table 5-33 Label on Rear Face Ref. Agilent Part Number Qty. Description Desig. 5188-1398 LABEL (Opt. 480) 5188-1399 LABEL (Opt. 485) 5185-3795 LABEL (Opt. 008) (Service Center Only) 5080-3939 LABEL (Opt. 010) (Service Center Only) 5185-3722 LABEL (Opt.
  • Page 260: Other Parts

    Replaceable Parts Replaceable Parts List Other Parts Table 5-34 Other Parts Agilent Part Number Qty. Description E5071-901x0 SERVICE GUIDE 1150-7799 MOUSE 5188-4407 KEYBOARD *1. The number indicated by “x” in the part number of each manual, sample program disk, or CD-ROM, 0 for the first edition, is incremented by 1 each time a revision is made.
  • Page 261: Power Cables And Plug Configurations

    Replaceable Parts Power Cables and Plug Configurations Power Cables and Plug Configurations Figure 5-34 Power Cables and Plug Configurations Chapter 5...
  • Page 262 Replaceable Parts Power Cables and Plug Configurations Chapter 5...
  • Page 263: Replacement Procedure

    Replacement Procedure This chapter provides procedure for removing and replacing the major assemblies in the E5071C ENA Series Network Analyzer.
  • Page 264: Replacing An Assembly

    Replacement Procedure Replacing an Assembly Replacing an Assembly The following steps show the sequence for replacing an assembly in a E5071C RF Network Analyzer. 1. Identify the faulty group. Refer to Chapter 4, “Troubleshooting.” 2. Order a replacement assembly. Refer to Chapter 5, “Replaceable Parts.”...
  • Page 265: Required Tools

    Replacement Procedure Required Tools Required Tools The following tools are required for repair of E5071C. Table 6-1 Required Tools Assembly TORX screwdriver flat edge torque cutting hex key open torque wrench open screwdriver screwdriver plier torque TORX T10 driver 1.5 mm 9/16 in. 5/8 in.
  • Page 266: Outer Cover Removal

    TORX screwdriver, T10, T15, and T20 Procedure Refer to Figure 6-1 for this procedure. Step 1. Disconnect the power cable from the E5071C. Step 2. Remove the four bottom feet (item 1). Step 3. Remove the four TORX T10 screws (item 2) from the bottom.
  • Page 267: Front Panel Removal

    Replacement Procedure Front Panel Removal Front Panel Removal Tools required • TORX screwdriver, T10, T15, and T20 • Flat edge screwdriver Procedure Refer to Figure 6-2 for this procedure. Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266. Step 2.
  • Page 268: Removable Hard Disk Assembly Removal

    Replacement Procedure Removable Hard Disk Assembly Removal Removable Hard Disk Assembly Removal Tools Required • TORX screwdriver, T10, T15, and T20 • Flat edge screwdriver Removal Procedure Refer to Figure 6-3 for this procedure. Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266. Step 2.
  • Page 269 Replacement Procedure Removable Hard Disk Assembly Removal Figure 6-3 Removable Hard Disk Assembly Replacement e5071cse1096 Chapter 6...
  • Page 270: Removable Hard Disk Drive Replacement

    Replacement Procedure Removable Hard Disk Drive Replacement Removable Hard Disk Drive Replacement Tools Required • TORX screwdriver, T10 • Torque screwdriver, TORX T10 (set to 0.29 N-m / 2.6 lb-in) Removal procedure Refer to Figure 6-4 for this procedure. Step 1. Insert the key (P/N 0960-2562) in the lock and turn clockwise (item 1). Step 2.
  • Page 271: Replacement Procedure

    Step 3. Turn the key to the left until it is locked, then remove the key. NOTE The “U” is displayed at the 8 segment LED on next to the removal hard disk when E5071C is turned on without key lock.
  • Page 272: Restore Backup Files Onto The New Hard Disk Drive

    Once the program has terminated, restart the E5071C to enable the system correction data. Calibration of the Touch Screen When you have replaced the HDD on a E5071C equipped, you have to calibrate the touch screen. Follow the procedure described below to calibrate the touch screen.
  • Page 273 Replacement Procedure Removable Hard Disk Drive Replacement Figure 6-5 Touch Panel Calibration Screen e5071cse1078 Step 5. Touch the x mark on the upper left with your finger. The mark x appears also on the lower left, upper right, and lower right. Touch the x marks in that order with your finger. Touching the four locations described above with your finger automatically concludes the touch screen calibration.
  • Page 274: Removable Hard Disk Assembly Removal (Option 017)

    Replacement Procedure Removable Hard Disk Assembly Removal (Option 017) Removable Hard Disk Assembly Removal (Option 017) Tools Required • TORX screwdriver, T10, T15, and T20 • Flat edge screwdriver Removal Procedure Refer to Figure 6-6 for this procedure. Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266. Step 2.
  • Page 275 Replacement Procedure Removable Hard Disk Assembly Removal (Option 017) Figure 6-6 Removable Hard Disk Assembly Replacement Opton 017 e5071cse3003 Chapter 6...
  • Page 276: Removable Hard Disk Drive Replacement Option 017

    Replacement Procedure Removable Hard Disk Drive Replacement Option 017 Removable Hard Disk Drive Replacement Option 017 Tools Required • TORX screwdriver, T10 • Torque screwdriver, TORX T10 (set to 0.50 N-m / 4.4 lb-in) Removal procedure Refer to Figure 6-7 for this procedure. Step 1.
  • Page 277: Replacement Procedure

    Replacement Procedure Removable Hard Disk Drive Replacement Option 017 Figure 6-7 Removable Hard Disk Drive Removal Option 017 e5071cse3004 Replacement Procedure Step 1. Replace the removable hard disk drive by inverse procedure of removal. NOTE Fasten the TORX T10 screws (item 7) using a torque screwdriver. Step 2.
  • Page 278: Hard Disk Assembly Removal Option 019

    Replacement Procedure Hard Disk Assembly Removal Option 019 Hard Disk Assembly Removal Option 019 Tools Required • TORX screwdriver, T10, T15, and T20 • Flat edge screwdriver Removal Procedure Refer to Figure 6-8 for this procedure. Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266. Step 2.
  • Page 279: Hard Disk Drive Replacement Option 019

    Replacement Procedure Hard Disk Drive Replacement Option 019 Hard Disk Drive Replacement Option 019 Tools Required • TORX screwdriver, T10 • Torque screwdriver, TORX T10 (set to 0.50 N-m / 4.4 lb-in) Removal procedure Refer to Figure 6-9 for this procedure. Step 1.
  • Page 280 Replacement Procedure Hard Disk Drive Replacement Option 019 Figure 6-9 Removable Hard Disk Drive Removal Option 019 e5071cse3006 Chapter 6...
  • Page 281: Replacement Procedure

    Replacement Procedure Hard Disk Drive Replacement Option 019 Replacement Procedure Step 1. Paste the heat transfer sheet to the hard disk drive rear side. Figure 6-10 Removable Hard Disk Drive Removal Option 019 Step 2. Replace the hard disk drive by inverse procedure of removal. NOTE Fasten the TORX T10 screws (item 3) using a torque driver.
  • Page 282: Power Supply Assembly Replacement

    Replacement Procedure Power Supply Assembly Replacement Power Supply Assembly Replacement Tools Required • TORX screwdriver, T10, T15, and T20 • Torque screwdriver, TORX T10 (set to 0.79 N-m / 7 lb-in) Removal Procedure Refer to Figure 6-11 for this procedure. Step 1.
  • Page 283: Replacement Procedure

    Replacement Procedure Power Supply Assembly Replacement Figure 6-11 Power Supply Assembly Removal e5071cse1098 Replacement Procedure Step 1. Replace the power supply assembly by inverse procedure of removal. NOTE Fasten the TORX T10 screws (item 7) using a torque screwdriver. Chapter 6...
  • Page 284: Digital Bridge Board Replacement

    Replacement Procedure Digital Bridge Board Replacement Digital Bridge Board Replacement Tools Required • TORX screwdriver, T10, T15, and T20 Removal Procedure Refer to Figure 6-12 for this procedure. Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266. Step 2.
  • Page 285: Replacement Procedure

    Replacement Procedure Digital Bridge Board Replacement Figure 6-12 Digital Bridge Board Removal e5071cse1099 Replacement Procedure Step 1. Replace the digital bridge board by inverse procedure of removal. Chapter 6...
  • Page 286: Handler I/O Board Replacement

    Replacement Procedure Handler I/O Board Replacement Handler I/O Board Replacement Tools Required • TORX screwdriver, T10, T15, and T20 Removal Procedure Refer to Figure 6-13 for this procedure. Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266. Step 2.
  • Page 287: Replacement Procedure

    Replacement Procedure Handler I/O Board Replacement Figure 6-13 Handler I/O Board Removal e5071cse1100 Replacement Procedure Step 1. Replace the handler I/O board by inverse procedure of removal. Chapter 6...
  • Page 288: Usb (Usbtmc) Interface Card Replacement

    Replacement Procedure USB (USBTMC) Interface Card Replacement USB (USBTMC) Interface Card Replacement Tools Required • TORX screwdriver, T10, T15, and T20 Removal Procedure Refer to Figure 6-14 for this procedure. Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266. Step 2.
  • Page 289: Replacement Procedure

    Replacement Procedure USB (USBTMC) Interface Card Replacement Figure 6-14 USB (USBTMC) Interface Card Removal e5071cse1102 Replacement Procedure Step 1. Replace the USB (USBTMC) interface card by inverse procedure of removal. Chapter 6...
  • Page 290: Pci Dsp Card And Gpib Card Replacement

    Replacement Procedure PCI DSP Card and GPIB Card Replacement PCI DSP Card and GPIB Card Replacement Tools Required • TORX screwdriver, T10, T15, and T20 • Open-end torque wrench, 5/8 inch (set to 1.97 N-m / 17.4 lb-in) Removal Procedure Refer to Figure 6-15 for this procedure.
  • Page 291: Replacement Procedure

    Replacement Procedure PCI DSP Card and GPIB Card Replacement Figure 6-15 PCI DSP Card and GPIB Card Removal e5071cse1101 Replacement Procedure Step 1. Replace the PCI DSP card and GPIB card by inverse procedure of removal. NOTE Fasten the 5/8 inch nuts (item 8) using an open-end torque wrench. Chapter 6...
  • Page 292: Digital Motherboard Replacement

    Replacement Procedure Digital Motherboard Replacement Digital Motherboard Replacement Tools Required • TORX screwdriver, T10, T15, and T20 Removal Procedure Refer to Figure 6-16 for this procedure. Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266 Step 2.
  • Page 293: Replacement Procedure

    Replacement Procedure Digital Motherboard Replacement Figure 6-16 Digital Motherboard Removal e5071cse1124 Replacement Procedure Step 1. Replace the digital motherboard by inverse procedure of removal. Chapter 6...
  • Page 294: Analog Base Module Replacement

    Replacement Procedure Analog Base Module Replacement Analog Base Module Replacement Tools Required • TORX screwdriver, T10, T15, and T20 • Open-end torque wrench, 5/8 inch (set to 1.97 N-m / 17.4 lb-in) Removal Procedure Refer to Figure 6-16 for this procedure. Step 1.
  • Page 295: Replacement Procedure

    Replacement Procedure Analog Base Module Replacement Figure 6-17 Analog Base Module Removal e5071cse1123 Replacement Procedure Step 1. Replace the analog base module by inverse procedure of removal. NOTE Fasten the 5/8 inch nuts (item 2) using an open-end torque wrench. Chapter 6...
  • Page 296: Level Vernier Module Replacement

    Replacement Procedure Level Vernier Module Replacement Level Vernier Module Replacement Tools Required • TORX screwdriver, T10, T15, and T20 • Flat edge screwdriver • Open-end wrench, 5/8 inch • Open-end torque wrench, 5/16 inch (set to 1.08 N-m / 9.5 lb-in) (for reconnecting SMA connector) •...
  • Page 297: Replacement Procedure

    Replacement Procedure Level Vernier Module Replacement Figure 6-18 Level Vernier Module Removal e5071cse1103 Replacement Procedure Step 1. Replace the level vernier module by inverse procedure of removal. NOTE Fasten the semi rigid cables using a torque wrench or torque driver set for replacement. Chapter 6...
  • Page 298: Synthesizer Module Replacement

    Replacement Procedure Synthesizer Module Replacement Synthesizer Module Replacement Tools Required • TORX screwdriver, T10, T15, and T20 • Flat edge screwdriver • Open-end wrench, 5/8 inch • Open-end torque wrench, 5/16 inch (set to 1.08 N-m / 9.5 lb-in) (for reconnecting SMA connector) •...
  • Page 299: Replacement Procedure

    Replacement Procedure Synthesizer Module Replacement Figure 6-19 Synthesizer Module Removal e5071cse1104 Replacement Procedure Step 1. Replace the synthesizer module by inverse procedure of removal. NOTE Fasten the semi rigid cables using a torque wrench or torque driver set for replacement. Chapter 6...
  • Page 300: Receiver Module Replacement For 2-Port Model

    Replacement Procedure Receiver Module Replacement for 2-Port model Receiver Module Replacement for 2-Port model Tools Required • TORX screwdriver, T10, T15, and T20 • Flat edge screwdriver • Open-end wrench, 5/8 inch • Open-end torque wrench, 5/16 inch (set to 1.08 N-m / 9.5 lb-in) (for reconnecting SMA connector) •...
  • Page 301: Replacement Procedure

    Replacement Procedure Receiver Module Replacement for 2-Port model Figure 6-20 Receiver Module Removal for 2-port model e5071cse1164 Replacement Procedure Step 1. Replace the receiver bodule by inverse procedure of removal. NOTE Fasten the semi rigid cables using a torque wrench or torque driver set for replacement. Chapter 6...
  • Page 302: Receiver Module Replacement For 4-Port Model

    Replacement Procedure Receiver Module Replacement for 4-Port model Receiver Module Replacement for 4-Port model Tools Required • TORX screwdriver, T10, T15, and T20 • Flat edge screwdriver • Open-end wrench, 5/8 inch • Open-end torque wrench, 5/16 inch (set to 1.08 N-m / 9.5 lb-in) (for reconnecting SMA connector) •...
  • Page 303: Replacement Procedure

    Replacement Procedure Receiver Module Replacement for 4-Port model Figure 6-21 Receiver Module Removal for 4-port model e5071cse1105 Replacement Procedure Step 1. Replace the receiver module by inverse procedure of removal. NOTE Fasten the semi rigid cables using a torque wrench or torque driver set for replacement. Chapter 6...
  • Page 304: Rf Switch And Distributor Module Replacement For 2-Port Model

    Replacement Procedure RF Switch and Distributor Module Replacement for 2-Port model RF Switch and Distributor Module Replacement for 2-Port model Tools Required • TORX screwdriver, T10, T15, and T20 • Flat edge screwdriver • Open-end wrench, 5/8 inch • Open-end torque wrench, 5/16 inch (set to 1.08 N-m / 9.5 lb-in) (for reconnecting SMA connector) •...
  • Page 305: Replacement Procedure

    Replacement Procedure RF Switch and Distributor Module Replacement for 2-Port model Figure 6-22 RF Switch and Distributor Module Replacement for 2-port model e5071cse1165 Replacement Procedure Step 1. Replace the switch and distributor module by inverse procedure of removal. NOTE Be careful screw (item 2) and screw (item 3) are different. NOTE Fasten the semi rigid cables, attenuator and terminations using a torque wrench or torque driver set for replacement.
  • Page 306: Rf Switch And Distributor Module Replacement For 4-Port Model

    Replacement Procedure RF Switch and Distributor Module Replacement for 4-Port model RF Switch and Distributor Module Replacement for 4-Port model Tools Required • TORX screwdriver, T10, T15, and T20 • Flat edge screwdriver • Open-end wrench, 5/8 inch • Open-end torque wrench, 5/16 inch (set to 1.08 N-m / 9.5 lb-in) (for reconnecting SMA connector) •...
  • Page 307: Replacement Procedure

    Replacement Procedure RF Switch and Distributor Module Replacement for 4-Port model Figure 6-23 RF Switch and Distributor Module Replacement for 4-port model e5071cse1106 Replacement Procedure Step 1. Replace the switch by inverse procedure of removal. NOTE Be careful screw (item 4) and screw (item 5) are different. NOTE Fasten the semi rigid cables using a torque wrench or torque driver set for replacement.
  • Page 308: Fan Replacement

    Replacement Procedure Fan Replacement Fan Replacement Tools Required • TORX screwdriver, T10, T15, and T20 • Flat edge screwdriver • A fine stick such as Hex key, 1.5 mm • Cutting plier or scissors Removal Procedure Refer to Figure 6-24 for this procedure. Step 1.
  • Page 309: Replacement Procedure

    Replacement Procedure Fan Replacement Figure 6-24 Fan Removal e5071cse1107 Replacement Procedure Step 1. Replace the fan by inverse procedure of removal. NOTE Connecting the cables (item 1) as described below. • Connect the short cable to “J5” connector. • Connect the middle cable to “J6” connector. •...
  • Page 310: Pca Dc Bias Board (Opt.xx5) Replacement

    Replacement Procedure PCA DC Bias Board (Opt.xx5) Replacement PCA DC Bias Board (Opt.xx5) Replacement Tools Required • TORX screwdriver, T10, T15, and T20 • Open-end torque wrench, 9/16 inch (set to 3.46 N-m / 30.5 lb-in) Removal Procedure Refer to Figure 6-24 for this procedure. Step 1.
  • Page 311: 10 Mhz Oven Board (Opt. 1E5) Replacement

    Replacement Procedure 10 MHz Oven Board (Opt. 1E5) Replacement 10 MHz Oven Board (Opt. 1E5) Replacement Tools Required • TORX screwdriver, T10, T15, and T20 Removal Procedure Refer to Figure 6-14 for this procedure. Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266. Step 2.
  • Page 312: Analog Motherboard Replacement

    Replacement Procedure Analog Motherboard Replacement Analog Motherboard Replacement Tools Required • TORX screwdriver, T10, T15, and T20 • Flat edge screwdriver • Open-end wrench, 5/8 inch • Open-end torque wrench, 5/16 inch (set to 1.08 N-m / 9.5 lb-in) (for reconnecting SMA connector) •...
  • Page 313: Replacement Procedure

    Replacement Procedure Analog Motherboard Replacement Figure 6-27 Analog Motherboard Removal e5071cse1110 Replacement Procedure Step 1. Replace the analog motherboard by inverse procedure of removal. Chapter 6...
  • Page 314: Front Panel Cover Removal

    Replacement Procedure Front Panel Cover Removal Front Panel Cover Removal Tools Required • TORX screwdriver, T10, T15, and T20 • Flat edge screwdriver Removal Procedure Refer to Figure 6-28 for this procedure. Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266. Step 2.
  • Page 315: Control Touch Panel Replacement

    Replacement Procedure Control Touch Panel Replacement Control Touch Panel Replacement Tools Required • TORX screwdriver, T10, T15, and T20 • Flat edge screwdriver Removal Procedure Refer to Figure 6-28 for this procedure. Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266. Step 2.
  • Page 316: Inverter Board Replacement

    Replacement Procedure Inverter Board Replacement Inverter Board Replacement Tools Required • TORX screwdriver, T8, T10, T15, and T20 • Flat edge screwdriver Removal Procedure Refer to Figure 6-30 for this procedure. Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266. Step 2.
  • Page 317: Replacement Procedure

    Replacement Procedure Inverter Board Replacement Figure 6-30 Inverter Board Removal e5071cse1112 Replacement Procedure Step 1. Replace the inverter assembly by inverse procedure of removal. NOTE Fasten the TORX T15 screws (item 2) using a torque screwdriver (set to 1.02 N-m / 9.0 lb-in).
  • Page 318: Lcd Replacement

    Replacement Procedure LCD Replacement LCD Replacement Tools Required • TORX screwdriver, T10, T15, and T20 • Flat edge screwdriver Removal Procedure Refer to Figure 6-31 for this procedure. Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266. Step 2.
  • Page 319: Replacement Procedure

    Replacement Procedure LCD Replacement Figure 6-31 LCD Removal e5071cse1114 Replacement Procedure Step 1. Replace the LCD by inverse procedure of removal. NOTE Fasten the TORX T15 screws (item 5) using a torque screwdriver (set to 1.02 N-m / 9.0 lb-in). Step 2.
  • Page 320: Front Panel Keyboard Replacement

    Replacement Procedure Front Panel Keyboard Replacement Front Panel Keyboard Replacement Tools Required • TORX screwdriver, T10, T15, and T20 • Flat edge screwdriver • Hex key, 1.5 mm Removal Procedure Refer to Figure 6-32 for this procedure. Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266. Step 2.
  • Page 321: Replacement Procedure

    Replacement Procedure Front Panel Keyboard Replacement Figure 6-32 Front Panel Keyboard Removal e5071cse1115 Replacement Procedure Step 1. Insert the whole jut of the rubber key into the hole on the new front panel keyboard. Step 2. Replace the front panel keyboard by inverse procedure of removal. NOTE Replace the keyboard rubber to new one, if it’s pulled the keyboard rubber from the front panel keyboard.
  • Page 322: Front Usb Board Replacement

    Replacement Procedure Front USB Board Replacement Front USB Board Replacement Tools Required • TORX screwdriver, T10, T15, and T20 • Flat edge screwdriver • Hex key, 1.5 mm Removal Procedure Refer to Figure 6-32 for this procedure. Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266. Step 2.
  • Page 323: Replacement Procedure

    Replacement Procedure Front USB Board Replacement Figure 6-33 A22 Front Panel Keyboard Removal e5071cse1116 Replacement Procedure Step 1. Replace the USB board by inverse procedure of removal. Chapter 6...
  • Page 324: Probe Power And Power Switch Board Replacement

    Replacement Procedure Probe Power and Power Switch Board Replacement Probe Power and Power Switch Board Replacement Tools Required • TORX screwdriver, T8, T10, T15, and T20 • Flat edge screwdriver • Cutting plier or scissors Removal Procedure Refer to Figure 6-32 for this procedure. Step 1.
  • Page 325: Replacement Procedure

    Replacement Procedure Probe Power and Power Switch Board Replacement Figure 6-34 Probe Power and Power Switch Board Removal e5071cse1117 Replacement Procedure Step 1. Replace the prove power board or switch board by inverse procedure of removal. Chapter 6...
  • Page 326 Replacement Procedure Probe Power and Power Switch Board Replacement Chapter 6...
  • Page 327: Post-Repair Procedures

    Post-Repair Procedures This chapter lists the procedures required to verify the E5071C operation after an assembly is replaced with a new one.
  • Page 328: Post-Repair Procedures

    Post-Repair Procedures Post-Repair Procedures Post-Repair Procedures Table 7-1 Post Repair Procedures lists the required procedures that must be performed after the replacement of an assembly or CPU. These are the recommended minimum procedures to ensure that the replacement is successfully completed. It is recommended to confirm the performance executing with all of Adjustment and performance test as final step.
  • Page 329 Post-Repair Procedures Post-Repair Procedures Table 7-1 Post-Repair Procedures Replaced Required Adjustments Verification Assembly or Correction Constants (CC) Part Synthesizer Perform the following required Perform “To Execute the diagnostics Test” on Module adjustment. page 187 - Synthesizer Gain - Source Output Power Perform the follwoing performance test.
  • Page 330 Post-Repair Procedures Post-Repair Procedures Table 7-1 Post-Repair Procedures Replaced Required Adjustments Verification Assembly or Correction Constants (CC) Part Distributer Perform the following required Perform “To Execute the diagnostics Test” on Module adjustments. page 187 - Source Output Power - Receiver Port Characteristics Perform the follwoing performance test.
  • Page 331 Post-Repair Procedures Post-Repair Procedures Table 7-1 Post-Repair Procedures Replaced Required Adjustments Verification Assembly or Correction Constants (CC) Part PCA CPI DSP No adjustment needed Inspect the Booting Process Card Perform “To Execute the diagnostics Test” on page 187 Perform “To Check the Video Output” on page Removable Calibration of the Touch Screen.
  • Page 332 Post-Repair Procedures Post-Repair Procedures Chapter 7...
  • Page 333: Manual Changes

    This appendix contains the information required to adapt this manual to versions or configurations of the E5071C manufactured earlier than the current printing date of this manual. The information in this manual applies directly to E5071C units with the serial number that is printed on the title page of this manual.
  • Page 334: Manual Changes

    Manual Changes Manual Changes Manual Changes To adapt this manual to your E5071C, refer to Table A-1 and Table A-2. Table A-1 Manual Changes by Serial Number Serial Prefix or Number Make Manual Changes Table A-2 Manual Changes by Firmware Version...
  • Page 335: System Recovery

    System Recovery This appendix describes how to recover the operating system (Windows XP) when the operating system has been damaged.
  • Page 336: System Recovery

    System Recovery System Recovery System Recovery By executing system recovery, you can return the system of the E5071C (the Windows operating system and the firmware) to the factory state (at the time of purchase CAUTION Strictly follow the steps described below. If you do any operation other than the following steps, the system may not be recovered.
  • Page 337: Procedure To Execute The Factory Recovery

    NOTE After Step 5, if the E5071C freezes with the cursor does not blink, turn the power off and remove unnecessary USB device. Then, replace the keyboard with a PS/2 keyboard to execute the process again.
  • Page 338 Figure B-2 Recover Factory Backup e5071cse1034 Step 8. When the recovery is finished, push the enter to select OK. The E5071C restart automatically. Step 9. After restart, the screen for initial registration appears. Execute initial registration. For information on the execution procedure, refer to “Initial Registration of E5071C” on page 357.
  • Page 339 System Recovery System Recovery CAUTION Never turn off the power during the system recovery because doing so may cause serious damage to the E5071C. Appendix B...
  • Page 340: Procedure To Create The User Backup Image

    This section describes how to create the user backup image. The C and D drives’ contents saved in this procedure are recalled when the user recovery function is executed. Step 1. Close the E5071C application by pressing > Service Menu > Exit.
  • Page 341 System Recovery System Recovery Step 6. Click button. Next e5071cse1126 Step 7. Select “The entire disk contents or individual partition,” then click button. Next e5071cse1127 Appendix B...
  • Page 342 System Recovery System Recovery Step 8. Select the C drive and D drive. e5071cse1128 Step 9. Select the destination folder and input a file name, then click button. It is Next recommended to save the file to a directory which has enough space like the E drive. Also, never save the contents of the drive to the C, D or F drive.
  • Page 343 System Recovery System Recovery Step 10. Select “Set the options manually” and click button. Next e5071cse1130 Step 11. Select “Backup priority” in the left field and “High” in the right field, then click Next button. e5071cse1131 Appendix B...
  • Page 344 System Recovery System Recovery Step 12. If necessary, input comments and click button. The comments can be omitted. Next e5071cse1132 Step 13. Click button. Proceed e5071cse1133 Appendix B...
  • Page 345 Step 15. The following screen appears when the user backup image was created. Click e5071cse1135 CAUTION Never turn off the power during creating the user backup image because doing so may cause serious damage to the E5071C. Appendix B...
  • Page 346: Procedure To Execute The User Recovery Function

    Returns the contents of the C and D drives to a user-specified state. To use this function, you must create the user backup image in advance. For more information, see the description “Procedure to create the user backup image” on page 340. Step 1. Close the E5071C application by pressing > Service Menu > Exit. System Step 2.
  • Page 347 System Recovery System Recovery Step 6. Click button. Next e5071cse1137 Step 7. Select the saved backup image file, then click button. Next e5071cse1138 Appendix B...
  • Page 348 System Recovery System Recovery Step 8. Select “Restore disks or partitions,” then click button. Next e5071cse1139 Step 9. Select the C drive, then click button. Next e5071cse1143 Appendix B...
  • Page 349 System Recovery System Recovery Step 10. Select the C drive to restore, then click button. Next e5071cse1144 Step 11. Select “Active” and click button. Next e5071cse1145 Appendix B...
  • Page 350 System Recovery System Recovery Step 12. Click button. Next e5071cse1146 Step 13. Select “Yes, I want to assign a logical drive letter to the restored partition” and click Next button. e5071cse1147 Appendix B...
  • Page 351 System Recovery System Recovery Step 14. Select “Yes, I want to restore another partition or hard disk drive,” then click button. Next e5071cse1148 Step 15. Select the D drive to restore, then click button. Next e5071cse1149 Appendix B...
  • Page 352 System Recovery System Recovery Step 16. Select the D drive to restore, then click t button. e5071cse1150 Step 17. Select “Logical” and click button. Next e5071cse1151 Appendix B...
  • Page 353 System Recovery System Recovery Step 18. Click button. Next e5071cse1152 Step 19. Select “Yes, I want to assign a logical drive letter to the restored partition,” then click Next button. e5071cse1153 Appendix B...
  • Page 354 System Recovery System Recovery Step 20. Select “No, I do not.” and click button. Next e5071cse1154 Step 21. Select “Set the options manually” and click button. Next e5071cse1155 Step 22. Select “Restoration Priority” in the left field and “High” in the right field, then click Next Appendix B...
  • Page 355 System Recovery System Recovery button. e5071cse1156 Step 23. Click button. Proceed e5071cse1157 Appendix B...
  • Page 356 Never turn off the power during the recovery of the user backup image because doing so may cause serious damage to the E5071C. Step 25. After boot up, perform “Initial Registration of E5071C” on page 357 and “Calibration of the Touch Screen” on page 359.
  • Page 357: Initial Registration Of E5071C

    System Recovery Initial Registration of E5071C Initial Registration of E5071C When you start up the E5071C at the first time, you need to perform the initial registration of the Windows XP operating system of the E5071C. NOTE You cannot use the front panel keys during the initial registration of the E5071C therefore connect the mouse and the keyboard before turning on the power.
  • Page 358 Step 4. In the next dialog box , input in the box. Then, click the button (Figure agena Name Next> B-6). Figure B-6 Windows XP Professional Setup dialog box (2/2) e5071cse1075 Step 5. After a while, the E5071C restarts automatically. Then, the measurement display appears. Appendix B...
  • Page 359: Calibration Of The Touch Screen

    System Recovery Calibration of the Touch Screen Calibration of the Touch Screen After E5071C measurement screen appears, you have to calibrate the touch screen. Follow the procedure described below to calibrate the touch screen. Step 1. Press Step 2. Press Service Menu Step 3.
  • Page 360 System Recovery Calibration of the Touch Screen Appendix B...
  • Page 361: Firmware Update

    Firmware Update This appendix describes how to update the E5071C firmware. When you want to update the E5071C firmware, refer to this appendix.
  • Page 362: Firmware Update

    Step 1. Prepare the latest E5071C firmware (USB mass storage device). Step 2. Connect the mouse and the keyboard to the E5071C rear panel. Then turn the E5071C on. Step 3. Connect the USB mass storage device into the front USB port of the E5071C.
  • Page 363 Firmware Update Firmware Update 4. Open dialog box apperas. Confirm that is displayed at the File name E5071.msi column. Select it, then press the button, and the Windows instoller will appear. Open Figure C-2 Firmware Open dialog box e5071cse1082 Step 5. After the installation is finished, the instrument will be restarted and the measurement view will appear.
  • Page 364 Firmware Update Firmware Update Appendix C...
  • Page 365: Power Requirement

    Power Requirement...
  • Page 366: Preparation For Power Supply

    Power Requirement Preparation for Power Supply Preparation for Power Supply Before turning on power to the equipment, be sure to verify the following: Power Requirements The E5071B requires the following power source: Requirements Voltage 90 to 132 VAC or 198 to 264 VAC Frequency 47 to 63 Hz Maximum power consumption...
  • Page 367 Power Requirement Preparation for Power Supply Figure D-1 Power cable options Appendix D...
  • Page 368: Turning The Power On And Off

    Power Requirement Turning the Power ON and OFF Turning the Power ON and OFF Perform the following steps to turn the power ON or OFF. Turning the power ON Step 1. If the standby switch ( ) in the lower-left part of the front panel is in the pressed down ) position, press the switch to the popped up position ( Step 2.
  • Page 369: Messages

    This appendix explains what these messages mean by listing them in alphabetical order. Messages showing the status of the E5071C are displayed in the lower-left area of the E5071C LCD screen. These messages include error messages that occur during the...
  • Page 370 Messages execution of GPIB commands and others that indicate the internal status of the equipment. Error messages are indicated following the character string “[Err]” and can be read out by a GPIB command. Other kinds of messages are indicated without the “[Err]” character string and cannot be read out by a GPIB command.
  • Page 371: Error Messages

    Messages Additional standard needed Error Messages An error message is displayed against a red background in the instrument message/warning area in the lower left part of the screen. Pushing a front panel key or executing :DISP:CCL command clears the error message. Errors caused by the operation of a front panel key simply appear on the display.
  • Page 372 Messages Command error -100 Command error A comprehensive syntax error has occurred showing that this instrument cannot detect a more detailed error. This code simply shows that a command error defined in 11.5.1.1.4, IEEE488.2 has occurred. -222 Data out of range A data element (not violating the standard) outside the range defined by this instrument has been received.
  • Page 373 Messages Failed to control signal generator This error occurs when the control of the ECal module fails at time of executing an ECal command, SENS:CORR:COLL:ECAL:SOLTn or SENS:CORR:OFFS:COLL:ECAL :xxxx. The failure results from the failure to connect the ECal module to the USB port, failure of the ECal module, etc.
  • Page 374 Messages Frequency offset enabled Frequency offset enabled This error occurs when the normal calibration (SENS:CORR:COLL:OPEN etc.) is performed under the condition of the frequency offset is enabled (SENS:OFFS ON). Frequency offset not enabled This error occurs when the scalar-mixer calibration (SENS:CORR:OFFS:COLL:OPEN etc.) is performed under the condition of the frequency offset is not enabled (SENS:OFFS OFF).
  • Page 375 Invalid suffix The suffix does not comply with the syntax defined in 7.7.3.2, IEEE488.2. Or it does not suit E5071C. Log sweep requires 2 octave minimum span The span of sweep range is not satisfied the requirement for logarithmic sweep. The sweep type is automatically changed to linear sweep when this error occurs.
  • Page 376 Messages Missing parameter -109 Missing parameter The number of parameters is less than that required for the command, or the parameter has not been entered. For example, the command SENS{1-6}:SWE:POIN requires one more parameter. Therefore, when a message “SENS1:SWE:POIN” is sent to a correct program message “SENS1:SWE:POIN 201”...
  • Page 377 Messages Out of memory -225 Out of memory Insufficient memory is available in this instrument to perform the required operation. -220 Parameter error When a parameter-related error other than Errors -221 through -229 occurs, that error is displayed. -108 Parameter not allowed The number of parameters exceeds that required for the command.
  • Page 378 Messages Port 4 receiver overload (For Options 413 and 414 only) If this error occurs, the stimulus signal output is automatically turned to off. Port 4 receiver overload (For Options 413 and 414 only) The input to Test Port 4 exceeds the maximum input level. The measurement value is not correct.
  • Page 379 Messages Query DEADLOCKED An error occurring when VBA is executed. -430 Query DEADLOCKED The state that generates a “DEADLOCKED” Query error (see 6.3.1.7, IEEE488.2). This error occurs when both input and output buffers have become full, preventing the instrument from continuing processing, for example. -400 Query error A comprehensive query error has occurred showing that this instrument cannot detect a...
  • Page 380 Messages Specified channel hidden been executed. This error also occurs when you try to save a Touchstone file with power sweep measurement specified or with the frequency offset function set to ON. Specified channel hidden This error occurs when an attempt is made to activate a channel not on display using the DISP:WIND:ACT command.
  • Page 381 Messages Syntax error -102 Syntax error A command or data type that is not recognized exists. -310 System error One of the errors designated as “system errors” in this instrument has occurred. Target value not found This error occurs when the target is not found during the marker search analysis after specifying the target and executing the CALC:MARK:FUNC:EXEC and CALC:FUNC:EXEC commands.
  • Page 382 Valid power meter is not connected. (refer to “Power Calibration” as described in E5071C Online Help) This error occurs when the E5071C and the power meter is not connected with USB/GPIB interface, the GPIB addresses do not match between the power meter and the E5071C’s power meter, or the power meter is turned off, during the measurement for power calibration data or for that in scalar-mixer calibration.
  • Page 383: Warning Message

    Messages Auto port extension not allowed Warning Message A warning message is displayed in the instrument message/Warning area in the lower left part of the display against a gray background. Pushing a front panel key or executing :DISP:CCL command clears the message. This message simply appears on the display, being not known to a remote environment such as a GPIB.
  • Page 384 Messages Sweep mode changed to stepped sweep Sweep mode changed to stepped sweep This message appears when you set the sweep type to power sweep, power slope function to enabled (ON, correction coefficient is other than zero), power calibration to ON, frequency offset to enabled, spurious avoidance of frequency offset to enabled, or external signal source control to enabled, while the sweep mode is set to swept mode or high speed swept mode.
  • Page 385 REGIONAL SALES AND SUPPORT OFFICES For more information about Agilent Technologies test and measurement products, applications, services, and for a current sales office listing, visit our web site: http://www.agilent.com/find/tmdir. You can also contact one of the following centers and ask for a test and measurement sales representative.

Table of Contents