2. Select the analyzer CW frequency:
a. F'or the HP 8719D, press (13.5) @.
b. F'or the HP 872OD, press 12oJ @.
c F'or the HP 8722D, press @ @J.
3. Write the frequency counter reading on the "Performance Test Record."
If the Instrument Fails This 'l&t
1. If the frequency measured is close to specillcation limits (either in or out of
2. If the analyzer fails by a signillcant margin, the master time base probably
needs ac@stment. In this case, refer to the "Frequency Accuracy
Adjustment" in Chapter 3.
3. See the "Source Troubleshooting" chapter for related troubleshooting
information.