CAUTION:
POWER
Error Number There is either a hardware failure in the source or you have
179
attempted to set the power level too high. The analyzer allows
the output power to be set higher or lower than the speciEed
available power range. However, these output powers may be
unleveled or unavailable. Check to see if the power level you
set is within specifications. If it is, refer to the "Source
Troubleshooting" chapter.
POWER METER INVALID
Error Number The power meter indicates an out-of-range condition. Check
1 1 6
test setup.
POWER METER NOT SETTLED
Error Number Sequential power meter readings are not consistent. Verify that
the equipment is set up correctly. If so, preset the instrument
118
and restart the operation.
PWR MTR: NOTON/CONNECTEDORWRONGADDR
Error Number The power meter cannot be accessed by the analyzer. Verify
117
that the power meter address and model number set in the
analyzer match the address and model number of the actual
power meter.
OUT
MAY
BE
UNLEVELED
Service Key Menus and Error Messages