Note
The analyzer may display the message CAUTION TEST PORT
with the procedure.

Set the potentiometers to the minimum levels.

4. Set the low band power potentiometer (A59 R4), full counter-clockwise.
5. Set the offset potentiometers (A59 R70, R71, RBO), to their full
counter-clockwise position.
The end stops on lO-turn potentiometers are indicated by a clicking sound.
6. Set the slope potentiometers (A59 R86, R.87, R88), to their full clockwise
position.
3-22 Adjustments and