CALIBRATION ABORTED
Error Number You have changed the active channel during a calibration so the
74
calibration in progress was terminated. Make sure the
appropriate channel is active and restart the calibration.
CALIBRATION REQUIRED
Error Number A calibration set could not be found that matched the current
63
stimulus state or measurement parameter. You will have to
perform a new calibration.
Error Number The disk is being accessed by the analyzer and is found to
203
contain an 1 (hierarchical file system) or llles nested within
subdirectories. The analyzer does not support 1. Replace the
disk medium with a LIF or DOS formatted disk that does not
contain files nested within subdirectories.
Error Number There is either a hardware failure in the source or you have
attempted to set the power level too high. The analyzer allows
179
the output power to be set higher or lower than the specified
available power range. However, these output powers may be
unleveled or unavailable. Check to see if the power level you
set is within specifications. If it is, refer to the
troubleshooting.