Correction Constants Retrieval Procedure

EEPROM Backup Disk
By using the current EEPROM backup disk, you can download the correction
constants data into the instrument EEPROMs
Insert the 'EEPROM Backup Disk" into the HP 8753E disk drive.
1.
2.
Make sure the A7 jumper/switch is in the Alter position.
3.
Pr
to
EEPROM data for the analyzer. On the factory shipped EEPROM backup
disk, the flIertame is FILEl.
4.
Press
the instrument EEPROMs.
5.
Perform "Option Numbers Correction Constant ('l&t 56)."
6.
Press w and verify that good data was transferred to EEPROM by
performing a simple measurement.
Move the A7 jumper/switch back to its Normal position when you are done
7.
working with the instrument.
and Correction Constants
to download the correction constants data into