CORRECTION CONSTANTS NOT STORED
Error Number A store operation to the EEPROM was not successful. You must
3
change the position of the jumper on the A7 CPU assembly.
Refer to the "A7 CC Jumper Position Procedure" in the
"Adjustments and Correction Constants" chapter.
CORRECTION TURNED OFF
Error Number Critical parameters in your current instrument state do not
66
match the parameters for the calibration set, therefore
correction has been turned off. The critical instrument state
parameters are sweep type, start frequency, frequency span,
and number of points
CURRENT PARAMETER NOT IN CAL SET
Error Number Correction is not valid for your selected measurement
64
parameter. Either change the measurement parameters or
perform a new calibration.
DEADLOCK
Error Number A fatal firmware error occurred before instrument preset
111
completed.