Post-Repair Procedures
The following tables list the additional service procedures which you must
perform to ensure that the instrument is working correctly, following the
replacement of an assembly. These procedures can be located in either
Chapter 2 or Chapter 3.
Replaced
Assembly
None
41 Front Panel
None
42 Front Panel
None
A52
Output
None
Load Firmware*
CC Retrieval
Option Number CC (Test 50)
Source Pretune CC (lkt 43)
IF Ampliller CC ('Ikst 47)
EEPROM Backup Disk
1 If you have an EEPROM backup disk available, you only need to perform the llrst five tests listed.
2 OnIy for instruments with Firmware revisions 7.xx and above.
Assembly Replacement and Post-Repair Procedures
Adjustments/
Correction Constants (Ch. 8)
Power Adjustments
Internal Test 12
Internal lkst 23
Internal Test 24
Internal lkst 0
Internal lkst 23
System Veriftcation
System Verification
Internal l&t 0
Check A8 test point voltages
Power Level Test
Dynamic Range !b3t
or
System Verikation