Magnitude and Phase Stability of Cables

1. To measure magnitude and phase stability, press the following keys on the
analyzer:
2.
Connect a short at the end of the cable and then wait for the analyzer to
average the measurement 64 times
3.
Hold the cable in a straight line, and press the following keys to normalize
the displayed traces:
4. MakeagradualOCPbendinthemid
measurement averaging by pressing
5.
To change the scale of the displayed traces, press:
6.
To mark the end of the cable's specified range, place a marker on the
highest speciiied frequency of the cable. Press: @iiZGQ (enter the speciiled
frequency) Lc/n.
the
.
System Verification and
Performance Tests