Error Messages
This section contains an alphabetical list of the error messages that pertain to
servicing the analyzer. The information in the list includes explanations of the
displayed messages and suggestion to help solve the problem.
Note
The error messages that pertain to measurement applications
are included
Guide.
ADDITIONALSTANDARDSNEEDED
Error Number Error correction for the selected calibration class cannot be
68
computed until you have measured all the necessary standards.
ADDRESSED TO TALK WITH NOTHING TO SAY
Error Number You have sent a read command to the analyzer (such as ENTER
716) without first requesting data with an appropriate output
31
command (such as OUTPDATA). The analyzer has no data in the
output queue to satisfy the request.
AIR FLOW RESTRICTED: CHECK FAN FILTER
Error Number Something is restricting the air flow into the analyzer. Check for
any debris and clean or replace the fan filter.
20
ANALOG INPUT OVERLOAD
Error Number The power level of the analog input is too high. Reduce the
60
power level of the analog input source.
1040
Service Key Menus and Error Messages
the
in