Agilent Technologies HP 8719D Service Manual page 268

Network analyzers
Table of Contents

Advertisement

Self Diagnose Softkey
You can access the self diagnosis function by pressing, &ZG) %lWf~ BEMU
SELF DIAGM3SE. This function examines, in order, the pass/fail status of ail
If a failure is detected, the routine displays the assembly or assemblies most
probably faulty and assigns a failure probability factor to each assembly.
The analyzer has up to 80 routines that test, verify, and adjust the instrument.
This section describes those tests.
Internal Tests
This group of tests runs without external connections or operator interaction.
All return a PASS or FAIL condition. All of these tests run on power-up and
preset except as noted.
ALL
0
13-16, and 20. Use the front panel knob to scroll through the tests and
see which failed. If all pass, the test displays a PASS status. Each test
in the subset retains its own test status.
PRESET.
1
ROM/RAM tests 2, 3, and 4; then tests 5 through 11, 14, 15, and 16. If
any of these tests fail, this test returns a FAIL status Use the front
panel knob to scroll through the tests and see which failed. If all pass,
this test displays a PASS status. Each test in the subset retains its own
test status This same subset is available over HP-IB as "TST?". It is not
performed upon remote preset.
2
ROM. Part of the ROM/RAM tests and cannot be run separately. Refer
to the "Digital Control Troubleshooting" chapter for more information.
Runs the following subset of internal tests: llrst, the
Service Key Menus and Error Messages
1 O-9

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 8720dHp 8722d

Table of Contents