OVERLOAD ON INPUTR, POWER REDUCED
Error Number You have exceeded approximately + 14 dBm at one of the test
5 7
ports. The RF output power is automatically reduced to
-85 dBm. The annotation PJ.l appears in the left margin of the
display to indicate that the power trip function has been
activated. When this occurs, reset the power to a lower level,
then toggle the SOURCE--FUR on f4VF softkey to switch on the
power again.
PARALLEL PORT NOT AVAILABLE FOR GPIO
Error Number You have defined the parallel port as COPY for sequencing in
the HP-IB menu. To access the parallel port for general purpose
165
PARALLEL PORT NOT AVAILABLE FOR COPY
Error Number You have defmed the parallel port as general purpose I/O (GPIO)
167
for sequencing. The delinition was made under the CLOCAL] key
menus. 'RI access the parallel port for copy, set the selection to
Service Key Menus and Error Messages