Front Panel Assembly
Tools Required
n
small slot screwdriver
n
ESD (electrostatic discharge) grounding wrist strap
n
n
Removal
1. Disconnect the power cord.
2. Remove the front bottom feet (item 1).
3. Remove all of the RF cables that are attached to the front panel (item 2).
4. Remove the trim strips (item 3) from the top and bottom edges of the front
frame by prying under the strip with a small slot screwdriver.
5. Remove the six screws (item 4) from the top and bottom edges of the frame.
6. Slide the front panel over the test port connectors.
free from the instrument.
Replacement
1. Reverse the order of the removal procedure.
Note
When reconnecting semi-rigid cables, it is recommended that the
connections be torqued to 10 in-lb.
Assembly Replacement and Post-Repair Procedures