Agilent Technologies HP 8719D Service Manual page 313

Network analyzers
Table of Contents

Advertisement

SWEEP TIME TOO FAST
Error Number The fractional-N and digital IF circuits have lost
12
synchronization. Refer to the HP 87'19D/ZO~/Z,ZD Network
TEST ABORTED
Error Number You have prematurely stopped a service test.
113
Error Number You have exceeded approximately + 14 dBm at one of the test
ports (or 0 dBm at the A or B sampler, Option 012 Only). When
57
this occurs, reduce the power to a lower level.
TROUBLE! CHECKSET-UP AND START OVER
Service Error Your equipment setup for the adjustment procedure in progress
Number 115 is not correct. Check the setup diagram and instructions in the
"Adjustments and Correction Constants" chapter. Start the
procedure again.
WRONG DISK FORMAT, INITIALIZE DISK
Error Number You have attempted to store, load, or read file titles, but your
disk format does not conform to the Logical Interchange Format
77
or writing to it.
1 O-54
Service Key Menus and Error Messages