Agilent Technologies HP 8719D Service Manual page 432

Network analyzers
Table of Contents

Advertisement

Assembly
Tools Required
n
Removal
1. Remove the rear panel (refer to "Rear Panel Assembly" in this chapter).
2. Disconnect W30 from the high stability frequency reference board (A26).
3. Remove the BNC connector nut and washer from the "10 MI-Iz PRECISION
REFERENCE" connector (item 1) on the rear panel.
4. Remove the screw (item 4) that attaches the lD5 assembly to the rear panel.
5. Remove the screw (item 2) that secures the high stability frequency
reference board (A26) to the bracket.
6. Slide the board out of the bracket. Be careful not to lose the plastic spacer
washer (item 3) that is on the BNC connector as the board is being removed.
Replacement
1. Reverse the order of the removal procedure.
Note
Before reinserting the high stability frequency reference board
(item 3) is on the BNC connector.
Assembly Replacement and Post-Repair Procedures

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 8720dHp 8722d

Table of Contents