Interpreting the Verification Results
The following Sgures show typical verifkation results, with dump graphics
activated, that could appear on a system verification printout. These printouts
compare the data from your measurement results with the traceable data and
corresponding uncertainty specifications Use these printouts to determine
whether your measured data falls within the total uncertainty lhnits at all
frequencies.
1. Upper limit points as defined by the total system uncertainty specifkations
2. Lower limit points as detied by the total system uncertainty speciWatioions
3. Data measured at the factory.
4. Results of magnitude measurement as measured in performance verilkation.
5. Correction is turned on.
6. Measurement parameter SII linear magnitude
7. Serial number of device
8. Device being measured : sys ver 1 = 20 dB attenuator.
232 System Verification and
8
STOP 13.510 000 Ooo GHz