Agilent Technologies HP 8719D Service Manual page 63

Network analyzers
Table of Contents

Advertisement

Figure 2-12. Aligning the Center Conductor
46. Enter l&t 30 (using step keys, entry keys, or front panel knob). Repeat
steps 38 through 42 with the 25Q mismatch airline. See the above figures
46. The printout of the measurements shows both a plot of the measurement,
when
corresponding data. The plot includes both the measured data trace and the
supplied factory data trace. The listing includes only the measured data. If
there is a failure at any frequency, an asterisk will be next to the measured
data and the out-of-speci&ation measured data on the plot will be blanked
out.
System Verification and
H o l d w i t h
t w o f i n g e r s
P O R T 1
H o l d h e r e
Figure 2-13. Ibrquing the
is active, and a list of the measured frequencies with
H o l d w i t h
o n e h a n d
P O R T 2