Trigger Mode - Operation At Window; Trigger Mode - Operation At Off - A&D Omniace III RA2300A Instruction Manual

Table of Contents

Advertisement

13.1.3. Trigger Mode – Operation at WINDOW
When the signal level enters into the pre-set range of the trigger source channel (i.e. IN) or goes out
of the range of the trigger source channel (i.e. OUT), a trigger is generated. The trigger setup can
be made for all analog channels. A trigger is generated upon satisfying of either of the trigger
settings.
Example: When CH8 is set to OUT
CH8
13.1.4. Trigger Mode – Operation at OFF
Trigger caused by a signal input into amp is not made. Only the manual and external triggers are
effective.
When printing is made, only the trigger made after the startup is effective. If the trigger
condition is satisfied before the start of recording, the status enters into the wait for next
NOTE
trigger condition.
RA2300A Recorder (7001754-R01)
Upper level
Lower level
13-3
13. Trigger Settings (Capturing Target Data to be Recorded)
Trigger
Trigger

Advertisement

Table of Contents
loading

Table of Contents