A&D Omniace III RA2300A Instruction Manual page 167

Table of Contents

Advertisement

Q 14: What is the "peak" format?
Answer One-point data is composed of both maximum and minimum values.
Maximum and minimum points are detected at the maximum rate of amp. This controls the
number of data and allows completely catching changes in a signal. The following data is
peak format. Extension of .FPP and .ID signifies the peak format.
*The file, when a peak file is saved once again, with .DRT extension is also peak format.
Details
Real-time record to recording paper also uses peak format. No data is lost by changing the
paper feeding speed.
Q 15: How to set the recorder for unmanned operation?
Answer The Timer function of System allows this measurement.
The following is setting examples.
Example 1: Recording from 12:00, Jan 7 to 10:00, Jan 8 in 2005
Example 2: Recording from 9:00 to 10:00 every day
Example 3: Recording for 10 minutes every three hours for a month from now on.
Case of Example 1
Start Timer
Start date 2005/12/05/ 12:00
End date 2005/12/06/ 10:00
(A recording interval is not used)
Case of Example 2
Start Timer
Start date **/**/**** 9:00
Start date **/**/**** 10:00
(A recording interval is not used)
Case of Example 3
Start Timer
Start date 2005/01/05/12:00 (specify the time after the current time)
Details
End date 2005/02/06/12:00 (specify the time at the finish time of a measurement)
Recording interval
Recording starting interval: 2 hours.
Recording operation interval: 10 minutes
The Start button is automatically pushed at the time specified in the start timer, and then a
recording operation is initiated. After starting, this operation works according to the specified
interval and ending time
Comment:
In combination with the Trigger button function, Trigger detection can be waited during
specified time.
- If the Trigger mode is turned off in Memory Recorder mode, the memory auto-save takes
effect during Stop. This allows automatically saving memory data based on the Stop time.
Caution:
Recording is not initiated with only Recording Interval Function ticked. To use interval,
make sure to configure the Start Timer and tick.