A&D Omniace III RA2300A Instruction Manual page 82

Table of Contents

Advertisement

10. HD Recorder (Recording data in the internal HDD)
(7) Data format
Setup item
Description
Peak
One data consist of two values: maximum and minimum values. Regardless of
recording speed, peak data is stored with the highest A/D conversion speed of amp.
It is possible to observe waveform having high-frequency element for long period of
time.
Sampling
One data consist of one instantaneous value. The data number per one piece of
data is half. Consequently, the recording speed and recording length will double.
(8) Waveform chart printing
Settings related to waveform chart printing are made.
a)
b)
a) Waveform ON/OFF
ON/OFF for waveform chart printing is made. Changes during operation can be made.
b) Same printing speed as recording speed
When the check box is ticked, the waveform chart printing speed is automatically set to the
speed the same as current recording speed. To separate the recording speed from the chart
feed speed, do not check this box.
The highest chart feed speed is 100 mm/s. If the recording speed is higher than this speed,
NOTE
the printing is made in the highest printing speed.
c) Chart feed speed setup
When the check box for "Same printing speed as recording speed" is not ticked, the chart
feed speed is separated from the data recording speed.