13.7. Trigger Settings for Event unit

Trigger settings for event Unit (Optional RA23-145) and event Box (Optional RA23-146) differs from
those of other amps. The settings are made in the following screen.
●Event Unit (Optional RA23-145), Event Box (Optional RA23-146)
Trigger condition
H
L
NO
RA2300A Recorder (7001754-R01)
The H condition is satisfied when the input voltage reaches +2.5V or higher.
The L condition is satisfied when the input voltage becomes +0.5 V or lower.
Exempting from trigger conditions
13. Trigger Settings (Capturing Target Data to be Recorded)
AND: When all trigger conditions that
have been set are satisfied, a
trigger is activated.
OR: Either of the trigger conditions is
satisfied, a trigger is activated.
Trigger settings are made when the check
box is ticked.
Trigger conditions for channels 1 to
8 are set.
H, L, NO can be set.
If NO is set the check box is not
displayed.
Volt (Voltage input)
13-11