Trigger Mode Off; Trigger Filter - A&D Omniace III RA2300A Instruction Manual

Table of Contents

Advertisement

13.4.4. Trigger Mode OFF

Trigger caused by a signal input into amp is not made. Only the manual and external triggers are
effective.
When printing is made, only the trigger made after the startup is effective. If the trigger
NOTE
condition is satisfied before the start of recording, the status enters into the wait for next
trigger condition.

13.5. Trigger Filter

13.5.1. Trigger Filter
After the trigger condition that is set is satisfied, the trigger is activated after this satisfied condition
continues for a specified period of time. This function is effective to eliminate noises whose pulse
length is relatively short.
In the peak including by HD recording, make calculation based on Recording Sampling
NOTE
Rate = 1 s.
RA2300A Recorder (7001754-R01)
13. Trigger Settings (Capturing Target Data to be Recorded)
Setting the trigger mode to OFF
13-9