6. Auto Setup (Automatically Setting Recording Conditions)

6.1. Function Overview

The auto setup function automatically sets recording conditions by
referencing the signals currently input. The following parameters are
automatically set.
● Auto range (for analog amps only)
● Auto sample
To execute this function, press the Auto button.
While recording is made, the auto setup function is not available.
TIPS
The time needed to validate the auto setup varies depending on
the signals that are input. As the auto setup function sets
recording conditions roughly, set conditions manually if precise
setup is required.

6.2. Auto Range

This function automatically sets the range of analog amps. The auto range function is available in the
following amps.
2CH High-Resolution DC Amp Unit
2CH FFT Amp Unit
2CH High-Speed DC Amp Unit
2CH AC Strain Amp Unit
2CH TC/DC Amp Unit
TC/DC Amp Unit
F/V Converter Unit
2CH Vibration and RMS Amp Unit
2CH DC Strain Amp Unit
2CH Zero Suppression Amp Unit
The auto range is executed from an amp screen detailed windows.
See
「 Chapter 7 Amp Unit
If there is no signals, the reference measurement value will the zero level, the adjusted value
TIPS
will be the highest sensitivity. If the high-sensitivity disabled setting in a DC amp is effective,
the range is adjusted within the low-sensitivity range.
Unit Name
Model
AP11-101
AP11-102
AP11-103
AP11-104/104A ACST
AP11-106/106A TCDC
AP11-107
AP11-108
AP11-109
AP11-110
AP11-111
6-2
RA2300A Recorder (7001754-R01)
Abbreviation
HRDC
HSDC
FFT
TDC
FV
RMS
DCST
HRZS