(4) Recording length
The length of recording in HD is indicated in data count and re
cording time. Setup can be made in the setup dialog box that i
s displayed when the button is pressed.
The recording time changes in conjunction with the
change of the data count, and vice versa. The
recording time can be determined by multiplying
data number by sampling rate.
File size = Acquisition information + Data type x Acquisition channel
When the recording time is set to Maximum, recording is made up to the size of free space of HD.
As a user does not need to take care of recording length, the Maximum is best choice when
recording is made with Start and Stop.
If free space is shortened during HD recording, operation stops with an error. In this case,
NOTE
shorter recording time than specified recording length of time will result.

(5) Recording operation

HD recording operation can be made with trigger detection.
Setup item
Recording
start
start button
Start trigger
Marking
upon
trigger
detection
Once
Repeated
(6) Filing mode
Setup item
Description
linear
Operation stops when specified number of data is
recorded.
loop
Overwriting for specified number of data is made.
(Overwriting is made from old data.)
The operation stops when the Stop button is pressed.
RA2300A Recorder (7001754-R01)
Contents
with
When the Start button is pressed, recording begins.
After the Start button is pressed, recording starts upon trigger activation.
Up to newest 128 trigger points can be memorized. The marking points
can be checked on the replay monitor.
Recording operation completes after one recording action.
HD recording repeats until the Stop button is pressed. However in the
following cases, operation will be invalid because one recording
automatically finishes.
The recording mode is "Ring."
The recording length is "Up to the size of free space."
10. HD Recorder (Recording data in the internal HDD)
10-5
ms
μs
Recording start
Data
Recording stop
Recording start
Data
Continues until forced stop