6.3. Auto Sampling

The auto sampling function sets the monitor displaying speed, chart paper feed speed, memory
sampling rate, and file recording rate based on the signals currently input.

6.3.1. Target Setup Conditions

The relation between settable conditions and mode are as follows. The monitor displaying speed is
always settable.
Monitor displaying speed
Chart feed speed
Memory sampling rate
File recording rate
X-Y sampling rate
In the X-Y Recorder mode, the auto sampling does not function. (For X-Y data sampling rate,
TIPS
make adjustment with seeing the monitor.)

6.3.2. Adjustment Range

Chart feed speed
100 mm/s (1 ms)
50 mm/s (2 ms)
20 mm/s (5 ms)
10 mm/s (10 ms)
5 mm/s (20 ms)
1 mm/s (100 ms)
100 mm/min (60 ms)
50 mm/min (120 ms)
20 mm/min (300 ms)
10 mm/min (600 ms)
5 mm/min (1.2 s)
1 mm/min (6 s)
 The values in parentheses in chart feed speed corresponds to the data sampling speed.
RA2300A Recorder (7001754-R01)
6. Auto Setup (Automatically Setting Recording Conditions)
Pen
Memory
Recorder
Available
Available
Available
-
-
Available
-
-
-
-
Memory sampling speed
1 μs
2 μs
5 μs
10 μs
20 μs
50 μs
100 μs
200 μs
500 μs
1 ms
2 ms
5 ms
6-3
Measurement Mode
HD
Available
Available
-
-
Available
Available
-
Available: Adjusted, -: Not adjusted
File recording speed
10 μs
100 μs
200 μs
500 μs
1 ms
2 ms
5 ms
10 ms
20 ms
50 ms
100 ms
200 ms
Multi