5. Input Signal Monitor (Observing Input Signals)
(7)
REC icon
Status of printing and recording is shown. The contents differ depending on the measurement mode.
For more information, see descriptions of measurement modes in Chapters 8 to 12.
(8)
TOP bar
(10) Temporary stop
(12) Monitoring speed
(13) Digital indication
(9)
Trigger sync
This button sets whether updating for input waveform monitoring is made through the trigger
detection or not. To monitor high-speed and periodic signals, this function is effective.
(10) Temporary stop
This button halts monitoring. Pressing the button again resumes monitoring.
(11) Monitor switching
This button switches input waveform monitoring mode. The monitor mode changes with button. The
mode for monitoring varies depending on the measurement mode. See the table below.
Monitor mode
Input Monitor
Chart Feed Speed
Memory Sampling Speed
HD Recording Speed
※ Monitor speed setting becomes disabled when settings other than the Input Monitor mode is set.
※ This button does not appear in the Pen Recorder and X-Y Recorder modes.
(12) Monitoring speed
When the monitor mode is set to Input Monitor, monitoring speed can be selected. Use the jog dial
or the window to set.
If the monitoring mode is set to other than the Input Monitor, the monitoring speed indicates a
fixed-recording speed. If the setting is changed, the fixed speed is changed.
(9) Trigger sync
(11) Monitor switching
Monitor Speed
Description
Fixed-speed for input monitor
Monitor displaying same speed as the
chart feed speed.
Monitor displaying same speed as the
memory sampling speed
Monitor display same speed as the
HD recording speed
Input Monitor
File Monitor
(16) Key lock
Memory
Enabled
Disabled Enabled Enabled
Enabled Disabled Enabled
Disabled Enabled Enabled
5-4
RA2300A Recorder (7001754-R01)
(14) Timer indication
(15) Time
Recorder mode
HD
Multi
Enabled Enabled