17. How to Use Optional Units

17.1.Connecting optional units

Please insert blank panel into empty slots when optional units are not connected in
WARNING
order to avoid electrical shock and to prevent damages to the recorder unit caused
by irruptions of extraneous substances
Please turn the power off and unplug power strip from the recorder unit before
CAUTION
connecting or disconnecting optional units. If optional units are connected or
disconnected while the power of the recorder unit is ON, it may damage the optional
unit and/or the recorder unit.
When exchanging units, do not touch internal parts, Internal parts may be damaged
CAUTION
if touched by static body. Do not touch anything except for the panel when
exchanging units. This may cause damages.
When connecting optional units, please be sure of top and bottom of the units and
CAUTION
insert units in accordance with the guideline specified in input slot area. Tighten the
screw with flat-head screwdriver after connecting optional units to the recorder unit.
Connecting optional units requires a flat head screwdriver (screwdriver's end must
be thinner than 0.65 mm).
Above
Not in use
Right side
RS-232C
17-2
EVENT
AC Bridge Excitation
RA2300A Recorder (7001754-R01)
REMOTE