19. Troubleshooting

19.1.Troubleshooting and Inspection

If the recorder does not operate normally or need to be repaired after following the
NOTE
instructions for abnormal conditions, please contact our sales offices listed at the end of
this manual.
Problem
No power supply.
Nothing
displayed
on the screen
The screen frozen
No reactions when
pushing any button
on the touch panel.
No recording
Not start recording
when pushing the
start button.
Cannot save data to
the selected media.
Not
detected
the
media.
Cannot specify and
control
operation
through
a
communication
interface.
Possible cause
The power switch is turned off.
The power cord is not properly
connected.
The fuse is blown.
The screen is automatically
turned off.
Abnormal condition occurs in
the system.
A recording operation is in
progress in Memory mode or
Multi-recorder mode. The start
LED or the copy LED is on.
There is no recording paper.
The printer block is opened.
The
temperature
of
thermal head is abnormally
high.
(The
environment
temperature
is
over
degrees centigrade.)
The start trigger is on.
Recording
is
set
as
external
synchronization
recording
There is no chart paper.
The media is not formatted.
The disk space is insufficient.
The media is write-protected.
The media is not correctly
formatted.
The media is damaged.
The
communication
parameters are not specified
correctly.
Instruction
Turn on the power switch
After the power switch is turned off,
connect the power cord correctly, and then
turn on the switch again.
The AC line input fuse for this unit cannot
be replaced by the customer because this
fuse is placed inside the main unit.
Please contact us if the fuse may be
blown.
Touching any button turns it on.
Turn off the power switch, and then turn it
on again
Push the stop button on the operation
panel to stop measuring, and then push
any button.
Insert recording paper.
Close the cover of the printer block.
the
Use the main body in a 0 to 40 degrees
environment.
40
Turn off the start trigger.
Recording is started only after inputting a
an
pulse
signal
to
Therefore, input the signal, and then push
the start button.
Load a chart paper.
Format the media.
Erase unwanted files or use new media.
Unprotect the media.
Format the media correctly.
-
Specify addresses and communication
parameters correctly.
19-2
a
remote
terminal.
RA2300A Recorder (7001754-R01)
Refer to
Chapter
3
Chapter
3
end of
this
manual
Chapter
16
Chapter
2
Chapter
2
Chapter
3
Chapter
3
Chapter
20
Chapter
10
Chapter
17
Chapter
3
Chapter
16
Chapter
16
Chapter
16
Chapter
16

Advertisement

Table of Contents
loading

Table of Contents