13. Trigger Settings (Capturing Target Data to be Recorded)

13.4.3. Trigger Mode WINDOW

When the signal level enters into the pre-set range of the trigger source channel (i.e. IN) or goes out
of the range of the trigger source channel (i.e. OUT), a trigger is generated. The trigger setup can be
made for all analog channels. A trigger is generated upon satisfying of either of the trigger settings.
Trigger setting
Pressing this portion switches the slope.
・Trigger detection at rising edge of a
signal.
・Trigger detection at falling edge of a
signal
・OFF
Setting the trigger mode to WINDOW
Setting the upper level
Touch the screen and turn the jog dial to
set the signal level.
Touching this portion makes jump to the first
channel.
Touching
increment one by one.
Setting the lower level
13-8