13. Trigger Settings (Capturing Target Data to be Recorded)
13.4.3. Trigger Mode WINDOW
When the signal level enters into the pre-set range of the trigger source channel (i.e. IN) or goes out
of the range of the trigger source channel (i.e. OUT), a trigger is generated. The trigger setup can be
made for all analog channels. A trigger is generated upon satisfying of either of the trigger settings.
Trigger setting
Pressing this portion switches the slope.
・Trigger detection at rising edge of a
signal.
・Trigger detection at falling edge of a
signal
・OFF
Setting the trigger mode to WINDOW
Setting the upper level
Touch the screen and turn the jog dial to
set the signal level.
▲
Touching this portion makes jump to the first
channel.
▲
Touching
increment one by one.
▲
Setting the lower level
13-8
RA2300A Recorder (7001754-R01)
this
portion
makes
channel