9 Performance Considerations

Overview

WARNING
Pulse oximetry readings and pulse signals can be affected by certain ambient
environmental conditions, OxiMax pulse oximetry sensor application errors, and
certain patient conditions. See the appropriate sections of the manual for specific
safety information.
Safety Information, page 9
Performance Considerations, page 97
Verify the performance of the OxiMax N-600x™ pulse oximeter by following the
procedures outlined in Modifying and Testing the Oximeter, page 103. A qualified
service technician should perform these procedures prior to initial installation in a
clinical setting.
Performance Considerations

Primary Considerations

Application issues and certain patient conditions can affect the measurements of the
OxiMax N-600x pulse oximeter and cause the loss of the pulse signal.
Oximetry Application Issues
Incorrect sensor application
Failure to cover the sensor with opaque material in high ambient light conditions
Patient Conditions
Dysfunctional hemoglobins
Poor peripheral perfusion
Excessive patient activity
Venous pulsations
Intravascular dyes, such as indocyanine green or methylene blue
Dark pigment or externally applied coloring agents (nail polish, dye, pigmented cream)
Defibrillation
Service Manual
To Purchase, Visit
Avobus.com
10030881 Rev. B 12-2008
or call
1-800-674-3655
97