Product Specifications
Condition
Normal
Normal
Single Fault
Normal
Reversed
Single Fault
Condition
Single Fault
Normal
Reversed
38
Table 11. Patient Applied and Patient Isolation Risk Current
Patient Applied Risk Current
AC Line Polarity
Neutral Line
Closed
Open
Closed
Closed
Open
Closed
Patient Isolation Risk Current
AC Line Polarity
Neutral Line
Closed
Closed
10030881 Rev. B 12-2008
Power Line Ground
Cable
Closed
Closed
Open
Closed
Closed
Open
Power Line Ground
Cable
Closed
Closed
IEC 60601-1
UL 60601-1
100 µA
500 µA
100 µA
500 µA
IEC 60601-1
UL 60601-1
5000 µA