Leakage Current - Nellcor OXIMAX N-600X Series Service Manual

Pulse oximeter with alarm management system
Hide thumbs Also See for OXIMAX N-600X Series:
Table of Contents

Advertisement

Leakage Current

The following tables display the maximum earth and enclosure leakage current
allowed, as well as patient leakage.
Condition
Normal
Single Fault
Normal
Single Fault
Condition
Normal
Single Fault
Normal
Single Fault
Service Manual
To Purchase, Visit
Avobus.com
Table 10. Earth and Enclosure Leakage Current Specifications
Earth Leakage Current
AC Polarity
Line Cord
Normal
Closed
Open
Closed
Reversed
Closed
Open
Closed
Enclosure Leakage Current
Neutral
AC Line Polarity
Line Cord
Normal
Closed
Open
Closed
Reversed
Closed
Open
Closed
or call
1-800-674-3655
Neutral
Line Cord
Closed
Closed
Open
Closed
Closed
Open
Power Line
Ground
Closed
Closed
Open
Closed
Closed
Open
10030881 Rev. B 12-2008
Safety Tests
IEC 60601-1
UL 60601-1
500 µA
300 µA
1000 µA
500 µA
300 µA
1000 µA
IEC 60601-1
UL 60601-1
100 µA
500 µA
100 µA
500 µA
37