Operation with a Live Subject
Patient monitoring involves connecting the Durasensor™ DS-100A pulse oximetry
sensor to a live subject for a qualitative test.
To test using a live subject:
1. Ensure the oximeter is connected to an AC power source.
2. Connect a DOC-10 pulse oximetry cable to the oximeter SpO
3. Connect the DS-100A pulse oximetry sensor to the DOC-10 pulse oximetry cable.
4. Attach the sensor to a live subject as recommended in the OxiMax pulse oximetry
sensor's Directions For Use.
5. Press the ON/STANDBY button to turn the oximeter on and verify the oximeter is
operating.
6. The oximeter should stabilize on the subject's physiological signal in about 15 to
30 seconds. Verify the oxygen saturation and pulse rate values are reasonable for
the subject.

Pulse Oximetry Functional Tests

These tests utilize the pulse oximetry functional tester (Nellcor model SRC-MAX) to
verify the performance of the oximeter.
Service Manual
Figure 82. SpO
Sensor Port
2
10030881 Rev. B 12-2008
Performance Tests
Sensor Port.
2
117