Rating Of Nurse Call Relay; Environmental Conditions; Operating; Transport And Storage - Nellcor OXIMAX N-600X Series Service Manual

Pulse oximeter with alarm management system
Hide thumbs Also See for OXIMAX N-600X Series:
Table of Contents

Advertisement

Product Specifications

Rating of Nurse Call Relay

Maximum Input Voltage
Load Current
Minimum Resistance
Ground Reference
Electrical Isolation

Environmental Conditions

Operating

Temperature
Altitude
Atmospheric Pressure 70 kPa to 106 kPa
Relative Humidity

Transport and Storage

Temperature
Altitude
Atmospheric Pressure 50 kPa to 106 kPa
Relative Humidity

Performance Specifications

Measurement Range

%SpO
Saturation Range
2
Pulse Rate Range
Perfusion Range
30
30 VAC or VDC (polarity is not important)
120 mA continuous (peak 300 mA @ 100 ms)
26.5 ohms to 50.5 ohms (40.5 ohms typical) during alarms
Isolated Ground
1500 Volts
5 ºC to 40 ºC (41 ºF to 104 ºF)
-390 m to 3,012 m
(-1,254 ft. to 9,882 ft.)
(20.6 in. Hg to 31.3 in. Hg)
15% to 95% non-condensing
Not in shipping container
-20 ºC to 60 ºC
(-4 ºF to 140 ºF)
-390 m to 5,574 m
(-1,254 ft. to 18,288 ft.)
(14.7 in. Hg to 31.3 in. Hg)
15% to 95% non-condensing
1% to 100%
20 to 250 beats per minute (bpm)
0.03% to 20%
10030881 Rev. B 12-2008
In shipping container
-20 ºC to 70 ºC
(-4 ºF to 158 ºF)
-390 m to 5,574 m
(-1,254 ft. to 18,288 ft.)
50 kPa to 106 kPa
(14.7 in. Hg to 31.3 in. Hg)
15% to 95% non-condensing
Service Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents