Product Specifications
Rating of Nurse Call Relay
Maximum Input Voltage
Load Current
Minimum Resistance
Ground Reference
Electrical Isolation
Environmental Conditions
Operating
Temperature
Altitude
Atmospheric Pressure 70 kPa to 106 kPa
Relative Humidity
Transport and Storage
Temperature
Altitude
Atmospheric Pressure 50 kPa to 106 kPa
Relative Humidity
Performance Specifications
Measurement Range
%SpO
Saturation Range
2
Pulse Rate Range
Perfusion Range
30
30 VAC or VDC (polarity is not important)
120 mA continuous (peak 300 mA @ 100 ms)
26.5 ohms to 50.5 ohms (40.5 ohms typical) during alarms
Isolated Ground
1500 Volts
5 ºC to 40 ºC (41 ºF to 104 ºF)
-390 m to 3,012 m
(-1,254 ft. to 9,882 ft.)
(20.6 in. Hg to 31.3 in. Hg)
15% to 95% non-condensing
Not in shipping container
-20 ºC to 60 ºC
(-4 ºF to 140 ºF)
-390 m to 5,574 m
(-1,254 ft. to 18,288 ft.)
(14.7 in. Hg to 31.3 in. Hg)
15% to 95% non-condensing
1% to 100%
20 to 250 beats per minute (bpm)
0.03% to 20%
10030881 Rev. B 12-2008
In shipping container
-20 ºC to 70 ºC
(-4 ºF to 158 ºF)
-390 m to 5,574 m
(-1,254 ft. to 18,288 ft.)
50 kPa to 106 kPa
(14.7 in. Hg to 31.3 in. Hg)
15% to 95% non-condensing
Service Manual