Nellcor OXIMAX N-600X Series Service Manual page 151

Pulse oximeter with alarm management system
Hide thumbs Also See for OXIMAX N-600X Series:
Table of Contents

Advertisement

Error Code
721
722
723
724
725
726
727
728
729
730
731
732
733
734
735
736
737
738
739
740
741
743
Service Manual
To Purchase, Visit
Avobus.com
Table 26. Error Code Listing for Non-correctable Failures
Category
Battery charger error - instance #2.
Battery charger error - instance #3.
Battery charger error - instance #4.
Battery charger error - instance #5.
Battery charger error - instance #6.
Battery
Battery charger error - instance #7.
Battery charger error - instance #8.
Battery charger error - instance #9.
Battery critical without AC power.
Battery fail without AC power.
EEPROM
EEPROM CRC Error - EEPROM cleared.
Vref error on DC.
Vref
Vref error on AC.
Temp
Overtemperature (> 58 C).
Battery I charge.
Battery
Battery I discharge.
Temp
Temperature low.
Hardware failsafe occurred, maybe overtemperature.
Failsafe
Hardware failsafe occurred, maybe overcurrent.
Hardware failsafe occurred, cause unknown.
Battery
Message RX CRC error.
Error in the shut down sequence. Occurs if the power is cycled off and on
Shutdown
too quickly.
10030881 Rev. B 12-2008
or call
1-800-674-3655
Prompts, Error Messages, and Error Codes
Description
151