Repairing the Oximeter
5. Remove the speaker by gently releasing the retaining tab and sliding the speaker
assembly from its mounting.
Caution
Avoid using excessive force when pressing the speaker retaining tab.
6. Set aside or discard the original speaker assembly, so it will not be mixed with its
replacement.

Replacing the Alarm Speaker

To replace the alarm speaker:
WARNING
Do not allow other metal objects to come into contact with the speaker; permanent
damage may occur.
Caution
Handle the speaker ONLY by the edges of the metal ring to avoid damage.
1. Pull the holding clip back, and insert the speaker into the top case.
2. Connect the speaker wire harness to J9 on the Main Board PCB.
3. Follow the steps outlined in Reassembling the Oximeter, page 169.