P e r f o r m a n c e V e r i f i c a t i o n

Introduction

Equipment Needed

N-550B
This section discusses the tests used to verify performance following repairs
or during routine maintenance. All tests can be performed without removing
the N-550B cover. All tests except the battery charge and battery performance
tests must be performed as the last operation before the N-550B is returned to
the user.
If the N-550B fails to perform as specified in any test, repairs must be made to
correct the problem before the N-550B is returned to the user.
Equipment
Digital multimeter (DMM)
®
Durasensor
oxygen transducer
(sensor)
O
M
oxygen transducer
XI
AX
Pulse oximeter tester
Safety analyzer
Pulse oximetry cable
Data interface cable
Stopwatch

Table 1: Equipment Needed

Fluke Model 87 or equivalent
DS-100A
MAX-A
SRC-MAX (must meet current EN
61326-1, FCC Part 15, and UL 1010-1,
or IEC 61010-1, or CSA 22.2 No.
1010-1 specifications)
Must meet current AAMI ESI/1993 &
IEC 60601-1/1998 specifications
DOC-10
EIA-232 cable (optional)
Manual or electronic
Description
7