Modifying and Testing the Oximeter
Safety Testing Standards
The OxiMax N-600x pulse oximeter safety tests are performed in accordance with
and meet the following standards. Also, see Product Safety Standards, page 32.
•
•
•
•
Applicable tests for these standards are listed below. Technicians must be familiar
with the standards applicable to their respective institution and country. Test
equipment and its application must comply with the applicable standards:
•
•
Note:
For testing the patient applied risk current, the leakage test lead from the test equipment
must be connected to the SpO
a male 9-pin "D" type connector with all pins shorted together.
During these tests, the oximeter displays the following message:
"RECONNECT/REPLACE SENSOR".
Battery Check
Note:
Replace the battery every 24 months.
Battery Power
To check battery power:
1. Disconnect the oximeter from AC power and verify the AC Power indicator shuts off.
2. Verify the oximeter is operating normally and the Low Battery indicator is off. If
the Low Battery indicator is lit, charge the battery.
3. Connect the oximeter to an AC power source and verify the AC Power indicator
turns on and the oximeter is operating normally.
Battery Charge
Note:
The battery charge procedure should be performed before oximeter repairs when possible.
Ensure the battery is fully charged by inspecting the Battery Fuel Gauge Indicator,
which should be completely full, and check the Low Battery Indicator, which should
be off. If the fuel gauge is not completely full, or if the Low Battery Indicator is lit,
charge the battery by connecting it to AC power. See Connecting to an AC Power
Source, page 50 and Using Battery Power, page 51.
106
To Purchase, Visit
Avobus.com
IEC 60601-1: 1988 + A1: 1991 + A2: 1995
EN 60601-1: 1990 + A11: 1993 + A12: 1993 + A13: 1996
UL 60601-1 1st edition
CSA C22.2 No. 601.1 M90
Ground Integrity, page 36, for test values.
Leakage Current, page 37, for test values.
2
10030881 Rev. B 12-2008
or call
1-800-674-3655
Sensor Port through the DOC-10 pulse oximetry cable using
Service Manual